Accuracy in target localization in stereotactic radiosurgery.

D R Choi1, Y C Ahn, D Y Kim, S J Huh, J I Lee.   

Abstract

The accuracy in target localization of CT, MRI, and digital angiography and the isocentric deviation of linear accelerator were investigated for stereotactic radiosurgery. Twenty-five slice images using CT and MRI were obtained out of geometrical phantom which was designed to produce exact 3-dimensional coordinates of several points within a 0.1 mm error range. These diagnostic images were transferred to a 3-D treatment planning system through ethernet. Measured 3-D coordinates of these images from the planning system were compared to known values by geometrical phantom. Anterior-posterior and lateral films were taken by digital angiography for measurement of spatial accuracy. The accuracy of gantry isocenter aligned by laser localizer was also measured. Lead ball was located at the isocenter of linear accelerator and x-ray films were taken by 4-MV photon beam with gantry angles of 0, 90, 180 and 270 degrees, Overall procedure from CT scanning to treatment was carried out using the geometrical phantom. The target accuracy was verified by high energy x-ray portal film. The accuracy of diagnostic machines were within 2.1 mm except MR-axial images. In case of linear accelerator, the deviation of isocenter was within 0.7 mm. Finally, the total isocentric deviation of overall procedure was 1.3 +/- 0.5 mm (using CT localization).

Mesh:

Year:  1997        PMID: 9136109     DOI: 10.1016/s0958-3947(96)00151-3

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  3 in total

1.  Clinical results of a pilot study on stereovision-guided stereotactic radiotherapy and intensity modulated radiotherapy.

Authors:  Shidong Li; Lawrence R Kleinberg; Daniele Rigamonti; Moody D Wharam; Abdul Rashid; Juan Jackson; David Djajaputra; Shenjen He; Tunisia Creasey; Theodore L DeWeese
Journal:  Technol Cancer Res Treat       Date:  2010-12

2.  Reliability of stereotactic coordinates of 1.5-tesla and 3-tesla MRI in radiosurgery and functional neurosurgery.

Authors:  Hae Yu Kim; Sun-Il Lee; Seong Jin Jin; Sung-Chul Jin; Jung Soo Kim; Kyoung Dong Jeon
Journal:  J Korean Neurosurg Soc       Date:  2014-03-31

3.  Six degrees of freedom CBCT-based positioning for intracranial targets treated with frameless stereotactic radiosurgery.

Authors:  Anees Dhabaan; Eduard Schreibmann; Arsalan Siddiqi; Eric Elder; Tim Fox; Tomi Ogunleye; Natia Esiashvili; Walter Curran; Ian Crocker; Hui-Kuo Shu
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.