Monitoring of left ventricular ejection fraction with a miniature, nonimaging nuclear detector: accuracy and reliability over time with special reference to blood labeling.

T B Lindhardt1, B Hesse, N Gadsbøll.   

Abstract

BACKGROUND: The purpose of this study was to determine the accuracy of determinations of left ventricular ejection fraction (LVEF) by a nonimaging miniature nuclear detector system (Cardioscint) and to evaluate the feasibility of long-term LVEF monitoring in patients admitted to the coronary care unit, with special reference to the blood-labeling technique. METHODS AND
RESULTS: Cardioscint LVEF values were compared with measurements of LVEF by conventional gamma camera radionuclide ventriculography in 33 patients with a wide range of LVEF values. In 21 of the 33 patients, long-term monitoring was carried out for 1 to 4 hours (mean 186 minutes), with three different kits: one for in vivo and two for in vitro red blood cell labeling. The stability of the labeling was assessed by determination of the activity of blood samples taken during the first 24 hours after blood labeling. The agreement between Cardioscint LVEF and gamma camera LVEF was good with automatic background correction (r = 0.82; regression equation y = 1.04x + 3.88) but poor with manual background correction (r = 0.50; y = 0.88x - 0.55). The agreement was highest in patients without wall motion abnormalities. The long-term monitoring showed no difference between morning and afternoon Cardioscint LVEF values. Short-lasting fluctuations in LVEFs greater than 10 EF units were observed in the majority of the patients. After 24 hours, the mean reduction in the physical decay-corrected count rate of the blood samples was most pronounced for the two in vitro blood-labeling kits (57% +/- 9% and 41% +/- 3%) and less for the in vivo blood-labeling kit (32% +/- 26%). This "biologic decay" had a marked influence on the Cardioscint monitoring results, demanding frequent background correction.
CONCLUSION: A fairly accurate estimate of LVEF can be obtained with the nonimaging Cardioscint system, and continuous bedside LVEF monitoring can proceed for hours with little inconvenience to the patients. Instability of the red blood cell labeling during long-term monitoring necessitates frequent background correction.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9115067     DOI: 10.1016/s1071-3581(97)90064-7

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  20 in total

1.  Acute myocardial dysfunction and recovery: a common occurrence after coronary bypass surgery.

Authors:  W M Breisblatt; K L Stein; C J Wolfe; W P Follansbee; J Capozzi; J M Armitage; R L Hardesty
Journal:  J Am Coll Cardiol       Date:  1990-05       Impact factor: 24.094

2.  Continuous monitoring of left ventricular function by an ambulatory radionuclide detector in patients with coronary artery disease.

Authors:  N Tamaki; T Yasuda; R H Moore; J B Gill; C A Boucher; A M Hutter; H K Gold; H W Strauss
Journal:  J Am Coll Cardiol       Date:  1988-09       Impact factor: 24.094

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  The application of a non-imaging nuclear probe system to the critically ill: laboratory validation and a clinical trial.

Authors:  J N Shephard; R Underwood; T W Evans
Journal:  Nucl Med Commun       Date:  1994-08       Impact factor: 1.690

5.  Kit-labeled technetium-99m red blood cells (Tc-99m-RBC's) for clinical cardiac chamber imaging.

Authors:  S M Larson; G W Hamilton; P Richards; J L Ritchie
Journal:  Eur J Nucl Med       Date:  1978

6.  Ambulatory scintigraphic assessment of transient changes in left ventricular function: a new method for detection of silent myocardial ischaemia.

Authors:  M Pfisterer; S Regenass; J Müller-Brand; F Burkart
Journal:  Eur Heart J       Date:  1988-12       Impact factor: 29.983

7.  Spontaneous changes in left ventricular function over the first 24 hours of acute myocardial infarction: implications for evaluating early therapeutic interventions.

Authors:  F J Wackers; H J Berger; M A Weinberg; B L Zaret
Journal:  Circulation       Date:  1982-10       Impact factor: 29.690

8.  Frequency of transient reductions in left ventricular ejection fraction at rest in coronary artery disease.

Authors:  M R Freeman; L de Yang; A Langer; B Foley; P W Armstrong
Journal:  Am J Cardiol       Date:  1994-07-15       Impact factor: 2.778

9.  Clinical evaluation of cardiac function by ambulatory ventricular scintigraphic monitoring (VEST): validation and study of the effects of nitroglycerin and nifedipine in patients with and without coronary artery disease.

Authors:  I H Mohiuddin; H Kambara; T Ohkusa; R Nohara; T Fudo; S Ono; N Tamaki; H Ohtani; Y Yonekura; C Kawai
Journal:  Am Heart J       Date:  1992-02       Impact factor: 4.749

10.  Rate of change of left ventricular ejection fraction during exercise is superior to the peak ejection fraction for predicting functionally significant coronary artery disease.

Authors:  B S Sridhara; S Bhattacharya; X J Liu; P Broadhurst; A Lahiri
Journal:  Br Heart J       Date:  1993-12
View more
  1 in total

1.  Pharmacological modulation of the ATP sensitive potassium channels during repeated coronary occlusions: no effect on myocardial ischaemia or function.

Authors:  T B Lindhardt; N Gadsbøll; H Kelbaek; K Saunamäki; J K Madsen; P Clemmensen; B Hesse; S Haunsø
Journal:  Heart       Date:  2004-04       Impact factor: 5.994

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.