Cardiac output and circulating blood volume analysis by pulse dye-densitometry.

T Iijima1, T Aoyagi, Y Iwao, J Masuda, M Fuse, N Kobayashi, H Sankawa.   

Abstract

OBJECTIVE: Pulse dye-densitometry (PDD) is a newly developed method for monitoring the indocyanine green (ICG) concentration in an artery with which cardiac output (CO) and circulating blood volume (CBV) can be determined. We evaluated its accuracy for clinical use.
METHODS: In 7 patients under general anesthesia, ICG-sensitive optical probes (805 and 890 nm) were attached to a finger. Following injection of ICG, the arterial concentration of dye was recorded optically by the non-invasive test instrument and sampled arterial blood ICG concentration was also measured photometrically for comparison. In order to validate the PDD analysis, CO was also measured by both the dye dilution cuvette method and by thermodilution in 8 patients scheduled for coronary artery bypass grafting. In 30 other patients, CBV assessed by PDD was compared with its value estimated from body size.
RESULTS: The blood dye concentration correlated well with the values obtained by PDD (r = 0.953, p < 0.01). Mean bias for the test PDD CO was +0.15 +/- 0.72 min l-1 (not significant (n.s.)) compared with the cuvette method while the mean bias of the thermodilution method vs the cuvette method was +0.79 +/- 0.84 min l-1 (p < 0.0001.). The average value of CBV obtained by PDD was 3.81 +/- 1.39 L compared with that estimated value, 3.72 +/- 0.77 L (n.s.).
CONCLUSIONS: CO determined by PDD agrees well with cuvette densitometry, and somewhat less well with CO by thermodilution. The new method, by not requiring a pulmonary arterial catheter, is less invasive than either older method, and yields in addition a value of CBV.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9112203     DOI: 10.1023/a:1007339924083

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  22 in total

1.  Estimation of hepatic blood flow with indocyanine green.

Authors:  C M LEEVY; C L MENDENHALL; W LESKO; M M HOWARD
Journal:  J Clin Invest       Date:  1962-05       Impact factor: 14.808

2.  History of blood gas analysis. VII. Pulse oximetry.

Authors:  J W Severinghaus; Y Honda
Journal:  J Clin Monit       Date:  1987-04

3.  Less-invasive cardiac output monitoring by earpiece densitometry.

Authors:  R C Grasberger; N S Yeston
Journal:  Crit Care Med       Date:  1986-06       Impact factor: 7.598

4.  Monitoring of cardiac output by thermodilution after open-heart surgery.

Authors:  F H Kohanna; J N Cunningham
Journal:  J Thorac Cardiovasc Surg       Date:  1977-03       Impact factor: 5.209

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Accuracy and reproducibility of the measurement of actively circulating blood volume with an integrated fiberoptic monitoring system.

Authors:  H Kisch; S Leucht; M Lichtwarck-Aschoff; U J Pfeiffer
Journal:  Crit Care Med       Date:  1995-05       Impact factor: 7.598

Review 7.  Pulmonary artery catheterization--at the crossroads?

Authors:  L A Eidelman; R Pizov; C L Sprung
Journal:  Crit Care Med       Date:  1994-04       Impact factor: 7.598

8.  A method for measurement of total circulating blood volume using indocyanine green.

Authors:  K Haneda; T Horiuchi
Journal:  Tohoku J Exp Med       Date:  1986-01       Impact factor: 1.848

9.  Impedance electrodes positioned on proximal portions of limbs quantify fluid compartments in dogs.

Authors:  M R Scheltinga; W S Helton; J Rounds; D O Jacobs; D W Wilmore
Journal:  J Appl Physiol (1985)       Date:  1991-05

10.  Poor predictive value of hematocrit and hemodynamic parameters for erythrocyte deficits after extensive elective vascular operations.

Authors:  P R Cordts; W W LaMorte; J B Fisher; C DelGuercio; J Niehoff; L E Pivacek; R C Dennis; H Siebens; A Georgio; C R Valeri
Journal:  Surg Gynecol Obstet       Date:  1992-09
View more
  32 in total

1.  Persistent splanchnic hyperemia during upright tilt in postural tachycardia syndrome.

Authors:  Julian M Stewart; Marvin S Medow; June L Glover; Leslie D Montgomery
Journal:  Am J Physiol Heart Circ Physiol       Date:  2005-09-02       Impact factor: 4.733

2.  Determination of the standard value of circulating blood volume during anesthesia using pulse dye-densitometry: a multicenter study in Japan.

Authors:  Takehiko Iijima; Hiroshi Ueyama; Yoshiyuki Oi; Isao Fukuda; Hironori Ishihara; Hikaru Kohase; Yoshifumi Kotake; Kaoru Koyama; Hideki Miyao; Naoki Kobayashi
Journal:  J Anesth       Date:  2005       Impact factor: 2.078

3.  Cardiac output measurement by pulse dye densitometry: comparison with pulmonary artery thermodilution in post-cardiac surgery patients.

Authors:  Martijn Kroon; A B Johan Groeneveld; Yvo M Smulders
Journal:  J Clin Monit Comput       Date:  2006-01-25       Impact factor: 2.502

4.  Changes in regional blood volume and blood flow during static handgrip.

Authors:  Julian M Stewart; Leslie D Montgomery; June L Glover; Marvin S Medow
Journal:  Am J Physiol Heart Circ Physiol       Date:  2006-08-25       Impact factor: 4.733

5.  Brain natriuretic peptide concentrations after aneurysmal subarachnoid hemorrhage: relationship with hypovolemia and hyponatremia.

Authors:  Sanne M Dorhout Mees; Reinier G Hoff; Gabriel J E Rinkel; Ale Algra; Walter M van den Bergh
Journal:  Neurocrit Care       Date:  2011-04       Impact factor: 3.210

6.  Splanchnic hyperemia and hypervolemia during Valsalva maneuver in postural tachycardia syndrome.

Authors:  Julian M Stewart; Marvin S Medow; Leslie D Montgomery; June L Glover; Mark M Millonas
Journal:  Am J Physiol Heart Circ Physiol       Date:  2005-06-17       Impact factor: 4.733

Review 7.  [Determinants of insensible fluid loss. Perspiration, protein shift and endothelial glycocalyx].

Authors:  M Jacob; D Chappell; K Hofmann-Kiefer; P Conzen; K Peter; M Rehm
Journal:  Anaesthesist       Date:  2007-08       Impact factor: 1.041

8.  Stability of cirrhotic systemic hemodynamics ensures sufficient splanchnic blood flow after living-donor liver transplantation in adult recipients with liver cirrhosis.

Authors:  Tomohide Hori; Shintaro Yagi; Taku Iida; Kentaro Taniguchi; Kentaro Yamagiwa; Chiduru Yamamoto; Takashi Hasegawa; Koichiro Yamakado; Takuma Kato; Kanako Saito; Linan Wang; Mie Torii; Yukinobu Hori; Kan Takeda; Kazuo Maruyama; Shinji Uemoto
Journal:  World J Gastroenterol       Date:  2007-11-28       Impact factor: 5.742

9.  Design of liver functional reserve monitor based on three-wavelength from red light to IR.

Authors:  Fuli Ye; Huimiao Zhan; Guilian Shi
Journal:  Technol Health Care       Date:  2018       Impact factor: 1.285

Review 10.  Systemic hemodynamics in advanced cirrhosis: Concerns during perioperative period of liver transplantation.

Authors:  Tomohide Hori; Yasuhiro Ogura; Yasuharu Onishi; Hideya Kamei; Nobuhiko Kurata; Motoshi Kainuma; Hideo Takahashi; Shogo Suzuki; Takashi Ichikawa; Shoko Mizuno; Tadashi Aoyama; Yuki Ishida; Takahiro Hirai; Tomoko Hayashi; Kazuko Hasegawa; Hiromu Takeichi; Atsunobu Ota; Yasuhiro Kodera; Hiroyuki Sugimoto; Taku Iida; Shintaro Yagi; Kentaro Taniguchi; Shinji Uemoto
Journal:  World J Hepatol       Date:  2016-09-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.