Three-dimensional imaging characteristics of the HEAD PENN-PET scanner.

J S Karp1, R Freifelder, M J Geagan, G Muehllehner, P E Kinahan, R M Lewitt, L Shao.   

Abstract

UNLABELLED: A volume-imaging PET scanner, without interplane septa, for brain imaging has been designed and built to achieve high performance, specifically in spatial resolution and sensitivity. The scanner is unique in its use of a single annular crystal of Nal(Tl), which allows a field of view (FOV) of 25.6 cm in both the transverse and axial directions. Data are reconstructed into an image matrix of 128(3) with (2 mm)3 voxels, using three-dimensional image reconstruction algorithms.
METHODS: Point-source measurements are performed to determine spatial resolution over the scanner FOV, and cylindrical phantom distributions are used to determine the sensitivity, scatter fraction and counting rate performance of the system. A three-dimensional brain phantom and 18F-FDG patient studies are used to evaluate image quality with three-dimensional reconstruction algorithms.
RESULTS: The system spatial resolution is measured to be 3.5 mm in both the transverse and axial directions, in the center of the FOV. The true sensitivity, using the standard NEMA phantom (6 liter), is 660 kcps/microCi/ml, after subtracting a scatter fraction of 34%. Due to deadtime effects, we measure a peak true counting rate, after scatter and randoms subtraction, of 100 kcps at 0.7 mCi for a smaller brain-sized (1.1 liter) phantom, and 70 kcps for a head-sized (2.5 liter) phantom at the same activity. A typical 18F-FDG clinical brain study requires only 2 mCi to achieve high statistics (100 million true events) with a scan time of 30 min.
CONCLUSION: The HEAD PENN-PET scanner is based on a cost-effective design using Nal(Tl) and has been shown to achieve high performance for brain studies and pediatric whole-body studies. As a full-time three-dimensional imaging scanner with a very large axial acceptance angle, high sensitivity is achieved. The system becomes counting-rate limited as the activity is increased, but we achieve high image quality with a small injected dose. This is a significant advantage for clinical imaging, particularly for pediatric patients.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9098216

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  4 in total

Review 1.  History and future technical innovation in positron emission tomography.

Authors:  Terry Jones; David Townsend
Journal:  J Med Imaging (Bellingham)       Date:  2017-03-31

2.  Detection of mesial temporal lobe hypoperfusion in patients with temporal lobe epilepsy by use of arterial spin labeled perfusion MR imaging.

Authors:  R L Wolf; D C Alsop; I Levy-Reis; P T Meyer; J A Maldjian; J Gonzalez-Atavales; J A French; A Alavi; J A Detre
Journal:  AJNR Am J Neuroradiol       Date:  2001-08       Impact factor: 3.825

Review 3.  Silicon photomultiplier signal readout and multiplexing techniques for positron emission tomography: a review.

Authors:  Haewook Park; Minseok Yi; Jae Sung Lee
Journal:  Biomed Eng Lett       Date:  2022-07-16

4.  Simulation of the expected performance of a seamless scanner for brain PET based on highly pixelated CdTe detectors.

Authors:  Ekaterina Mikhaylova; Gianluca De Lorenzo; Mokhtar Chmeissani; Machiel Kolstein; Mario Cañadas; Pedro Arce; Yonatan Calderón; Dilber Uzun; Gerard Ariño; José Gabriel Macias-Montero; Ricardo Martinez; Carles Puigdengoles; Enric Cabruja
Journal:  IEEE Trans Med Imaging       Date:  2013-10-04       Impact factor: 10.048

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.