Literature DB >> 9089594

An investigation of tomotherapy beam delivery.

J N Yang1, T R Mackie, P Reckwerdt, J O Deasy, B R Thomadsen.   

Abstract

Experimental simulations for tomotherapy beam delivery were performed using a computer-controlled phantom positioner, a cylindrical phantom, and a 6 MV x-ray slit beam. Both continuous helical beam and sequential segmented tomotherapy (SST) beam deliveries were evaluated. Beam junctioning problem due to couch indexing error or field width errors presented severe dose uniformity perturbations for SST, while the problem was minimized for helical beam delivery. Longitudinal breathing motions were experimentally simulated for helical and SST beam delivery. While motions reduced the dose uniformity perturbations for SST, small artifacts in dose uniformity can be introduced for helical beam delivery. With typical breath frequency and magnitude, for a slit beam of 2.0 cm width at 4 rpm, the dose uniformity perturbation was not significant. A running start/stop technique was implemented with helical beam delivery to sharpen the 20%-80% longitudinal dose fall-off from 1.5 to 0.5 cm. The latter was comparable to the corresponding dose penumbra of a conventional 6 MV 10 x 10 cm2 field. All together, helical beam delivery showed advantages over SST for tomotherapy beam delivery under similar delivery conditions.

Entities:  

Mesh:

Year:  1997        PMID: 9089594     DOI: 10.1118/1.597909

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  Feasibility of intrafraction whole-body motion tracking for total marrow irradiation.

Authors:  Manju Sharma; Troy Dos Santos; Nikolaos P Papanikolopoulos; Susanta Kumar Hui
Journal:  J Biomed Opt       Date:  2011-05       Impact factor: 3.170

2.  On the making of sharp longitudinal dose profiles with helical tomotherapy.

Authors:  Michael W Kissick; Ryan T Flynn; David C Westerly; Thomas Rockwell Mackie; Peter W Hoban
Journal:  Phys Med Biol       Date:  2007-10-16       Impact factor: 3.609

3.  Whole brain radiotherapy with hippocampal avoidance and simultaneously integrated brain metastases boost: a planning study.

Authors:  Alonso N Gutiérrez; David C Westerly; Wolfgang A Tomé; Hazim A Jaradat; Thomas R Mackie; Søren M Bentzen; Deepak Khuntia; Minesh P Mehta
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-10-01       Impact factor: 7.038

Review 4.  The radiation techniques of tomotherapy & intensity-modulated radiation therapy applied to lung cancer.

Authors:  Zhengfei Zhu; Xiaolong Fu
Journal:  Transl Lung Cancer Res       Date:  2015-06

5.  A new method to reconstruct intra-fractional prostate motion in volumetric modulated arc therapy.

Authors:  Y Chi; N H Rezaeian; C Shen; Y Zhou; W Lu; M Yang; R Hannan; X Jia
Journal:  Phys Med Biol       Date:  2017-07-07       Impact factor: 3.609

6.  Evaluation of tumor motion effects on dose distribution for hypofractionated intensity-modulated radiotherapy of non-small-cell lung cancer.

Authors:  Hyejoo Kang; Ellen D Yorke; Jie Yang; Chen-Shou Chui; Kenneth E Rosenzweig; Howard I Amols
Journal:  J Appl Clin Med Phys       Date:  2010-06-08       Impact factor: 2.102

7.  Treatment planning to improve delivery accuracy and patient throughput in helical tomotherapy.

Authors:  David C Westerly; Emilie Soisson; Quan Chen; Katherine Woch; Leah Schubert; Gustavo Olivera; Thomas R Mackie
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-04-23       Impact factor: 7.038

8.  Synchronized moving aperture radiation therapy (SMART): superimposing tumor motion on IMRT MLC leaf sequences under realistic delivery conditions.

Authors:  Jun Xu; Nikos Papanikolaou; Chengyu Shi; Steve B Jiang
Journal:  Phys Med Biol       Date:  2009-07-30       Impact factor: 3.609

9.  Feasibility of concurrent treatment with the scanning ultrasound reflector linear array system (SURLAS) and the helical tomotherapy system.

Authors:  José A Peñagarícano; Eduardo Moros; Petr Novák; Yulong Yan; Peter Corry
Journal:  Int J Hyperthermia       Date:  2008-08       Impact factor: 3.914

10.  Verification of tomotherapy dose delivery.

Authors:  S M Pelagade; B R Paliwal
Journal:  J Med Phys       Date:  2009-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.