Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 +/- 0.79%, 1.91 +/- 1.39%, 2.14 +/- 1.3%, 1.3 +/- 0.73% and 1.67 +/- 1.5% for head and neck, prostate, pelvis-abdomen sites, and for all other sites. The spatial agreement between the calculated and the measured film dose distributions was evaluated using the gamma metric distribution. The average frequency versus gamma interval was plotted as a bar graph to quantify the gamma index variation inside the region of interest for each body site.Entities:
Keywords: Dose; gamma index; tomotherapy
Year: 2009 PMID: 20098569 PMCID: PMC2807687 DOI: 10.4103/0971-6203.54856
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1The left hand side image shows 3D rendering of the phantom. The right hand side image shows cross-section view that exposes the location of the ion chamber
Figure 2% Dose difference (absolute) variation for all 71 Cases
Figure 3Gamma Analysis: On Y-axis average frequency represents the average dose pixels falling in one of the eight gamma index interval. The bar graph provides the average frequency value (for all 10 patients of each site) for various gamma intervals 0-0.25, 0.25-0.50, 0.50-0.75, 0.75-1.00, 1.00-1.25 and 1.25-1.50 from frequency gamma histogram