Literature DB >> 9084360

Energy cost of walking: comparison of "intelligent prosthesis" with conventional mechanism.

J G Buckley1, W D Spence, S E Solomonidis.   

Abstract

OBJECTIVE: To determine physiological energy cost with Blatchford's "Intelligent Prosthesis" (IP) compared to energy cost with a conventional pneumatic swing phase control (PSPC) mechanism.
DESIGN: Before-After trial: subjects fitted with IP walked on programmable treadmill at speeds: 6 min slow, 6 min fast, 8 min while speed changed, between slow, normal, and fast, every minute, and 6 min normal. Breath-by-breath analysis of subject's expired air determined average Vo2 (L/min) within each period. Procedure repeated after 1-week interval using PSPC prosthesis. Testing sessions supervised by experienced prosthetist.
SETTING: Rehabilitation centre.
SUBJECTS: Volunteer sample. Three men, unilateral transfemoral traumatic amputee patients, ages 39 to 59 years. Normally used ischial containment socket, Blatchford Endolite Stabilised Stance Flex knee with PSPC and Multiflex foot and ankle.
INTERVENTIONS: Fitting, programming, and alignment of IP (own socket) by Bioengineering Unit's resident prosthetist, IP's microprocessor programmed to facilitate five walking speeds. MAIN OUTCOME MEASURE: Physiological energy cost (Vo2), of using IP compared to using PSPC mechanism.
RESULTS: Two subjects displayed reduced Vo2 of between 5.6% and 9.0% using IP compared to PSPC prosthesis at a pace either faster or slower than their normal pace. Third subject showed no significant change in oxygen consumption despite IP unit being heavier. All subjects displayed reduced Vo2 (averaging 4.1%) using IP for period of variable speed walking.
CONCLUSIONS: Although differences were small, they tend to indicate that use of the heavier IP unit lowered the energy cost of walking at speeds other than the amputee's normal pace.

Entities:  

Mesh:

Year:  1997        PMID: 9084360     DOI: 10.1016/s0003-9993(97)90044-7

Source DB:  PubMed          Journal:  Arch Phys Med Rehabil        ISSN: 0003-9993            Impact factor:   3.966


  12 in total

Review 1.  Science, medicine, and the future: Artificial limbs.

Authors:  L J Marks; J W Michael
Journal:  BMJ       Date:  2001-09-29

2.  Transfemoral amputations: is there an effect of residual limb length and orientation on energy expenditure?

Authors:  Johanna C Bell; Erik J Wolf; Barri L Schnall; John E Tis; Benjamin K Potter
Journal:  Clin Orthop Relat Res       Date:  2014-10       Impact factor: 4.176

3.  Restoration of stance phase knee flexion during walking after spinal cord injury using a variable impedance orthosis.

Authors:  Thomas C Bulea; Rudi Kobetic; Ronald J Triolo
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2011

4.  Energy expenditure and activity of transfemoral amputees using mechanical and microprocessor-controlled prosthetic knees.

Authors:  Kenton R Kaufman; James A Levine; Robert H Brey; Shelly K McCrady; Denny J Padgett; Michael J Joyner
Journal:  Arch Phys Med Rehabil       Date:  2008-07       Impact factor: 3.966

5.  Powered knee and ankle prosthesis with indirect volitional swing control enables level-ground walking and crossing over obstacles.

Authors:  Joel Mendez; Sarah Hood; Andy Gunnel; Tommaso Lenzi
Journal:  Sci Robot       Date:  2020-07-22

6.  Finite state control of a variable impedance hybrid neuroprosthesis for locomotion after paralysis.

Authors:  Thomas C Bulea; Rudi Kobetic; Musa L Audu; John R Schnellenberger; Ronald J Triolo
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2012-11-15       Impact factor: 3.802

Review 7.  Measures and procedures utilized to determine the added value of microprocessor-controlled prosthetic knee joints: a systematic review.

Authors:  Patrick J R Theeven; Bea Hemmen; Peter R G Brink; Rob J E M Smeets; Henk A M Seelen
Journal:  BMC Musculoskelet Disord       Date:  2013-11-27       Impact factor: 2.362

8.  Design and Control of a New Biomimetic Transfemoral Knee Prosthesis Using an Echo-Control Scheme.

Authors:  Mario G Bernal-Torres; Hugo I Medellín-Castillo; Juan C Arellano-González
Journal:  J Healthc Eng       Date:  2018-04-30       Impact factor: 2.682

9.  Interactive cueing with Walk-Mate for hemiparetic stroke rehabilitation.

Authors:  Takeshi Muto; Barbara Herzberger; Joachim Hermsdoerfer; Yoshihiro Miyake; Ernst Poeppel
Journal:  J Neuroeng Rehabil       Date:  2012-08-21       Impact factor: 4.262

10.  Impact on gait biomechanics of using an active variable impedance prosthetic knee.

Authors:  Matthew R Williams; Susan D'Andrea; Hugh M Herr
Journal:  J Neuroeng Rehabil       Date:  2016-06-10       Impact factor: 4.262