PURPOSE: The purpose of this study was to investigate the influence of different parameters (object size, shape, orientation, contrast, observer, and window setting) on the accuracy of volume measurements in spiral CT.
METHOD: The phantom study consisted of two parts. First, well-circumscribed ellipsoid objects were scanned with conventional and spiral CT. Volumes were determined by two observers using manual contour delineation and summation-of-areas. The influence of object size, CT technique, and observer was assessed. Second, irregularly shaped gingerroots were scanned and the effect of observer, contrast, and orientation was investigated. Also assessed in both parts of the study was the effect of window setting.
RESULTS: Spiral CT offered higher accuracy and reproducibility than conventional CT did. The accuracy of volume measurements was mainly determined by the window setting used for image display. Using an "optimized" window center value halfway between object density and background density resulted in overestimation of true object volumes if the number of slices through the object (number of samples) was low. Other factors (contrast, observer-related errors in contour delineation) were less important.
CONCLUSIONS: Volume measurements of small objects, obtained with spiral CT and using manual contour delineation, may be grossly erroneous. Major causes of inaccuracy are an inappropriate window center selection and undersampling.

Mesh:

Year:  1997        PMID: 9071313     DOI: 10.1097/00004728-199703000-00034

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  8 in total

1.  Interobserver variability in gross tumor volume delineation for hepatocellular carcinoma : Results of Korean Radiation Oncology Group 1207 study.

Authors:  Young Suk Kim; Jun Won Kim; Won Sup Yoon; Min Kyu Kang; Ik Jae Lee; Tae Hyun Kim; Jin Hee Kim; Hyung-Sik Lee; Hee Chul Park; Hong Seok Jang; Chul Seung Kay; Sang Min Yoon; Mi-Sook Kim; Jinsil Seong
Journal:  Strahlenther Onkol       Date:  2016-08-18       Impact factor: 3.621

2.  Reliability of CT-based tumor volumetry after intraarterial chemotherapy in patients with small carcinoma of the oral cavity and the oropharynx.

Authors:  Stefan Rohde; Adorján F Kovács; Joachim Berkefeld; Bernd Turowski
Journal:  Neuroradiology       Date:  2006-04-12       Impact factor: 2.804

3.  Validation of multi-detector computed tomography as a non-invasive method for measuring ovarian volume in macaques (Macaca fascicularis).

Authors:  Jeryl C Jones; Susan E Appt; Stephen R Werre; Joshua C Tan; Jay R Kaplan
Journal:  Am J Primatol       Date:  2010-06       Impact factor: 2.371

4.  Volume assessment accuracy in computed tomography: a phantom study.

Authors:  Nicolas D Prionas; Shonket Ray; John M Boone
Journal:  J Appl Clin Med Phys       Date:  2010-04-16       Impact factor: 2.102

5.  Preclinical multimodality phantom design for quality assurance of tumor size measurement.

Authors:  Yongsook C Lee; Gary D Fullerton; Cristel Baiu; Margaret G Lescrenier; Beth A Goins
Journal:  BMC Med Phys       Date:  2011-09-30

6.  Phantom investigation of 3D motion-dependent volume aliasing during CT simulation for radiation therapy planning.

Authors:  James A Tanyi; Martin Fuss; Vladimir Varchena; Jack L Lancaster; Bill J Salter
Journal:  Radiat Oncol       Date:  2007-02-24       Impact factor: 3.481

7.  Evaluation of tomotherapy MVCT image enhancement program for tumor volume delineation.

Authors:  Spencer Martin; George Rodrigues; Quan Chen; Simon Pavamani; Nancy Read; Belal Ahmad; Alex Hammond; Varagur Venkatesan; James Renaud; Slav Yartsev
Journal:  J Appl Clin Med Phys       Date:  2011-04-12       Impact factor: 2.102

8.  Volumetric accuracy of cone-beam computed tomography.

Authors:  Cheol-Woo Park; Jin-Ho Kim; Yu-Kyeong Seo; Sae-Rom Lee; Ju-Hee Kang; Song-Hee Oh; Gyu-Tae Kim; Yong-Suk Choi; Eui-Hwan Hwang
Journal:  Imaging Sci Dent       Date:  2017-09-21
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.