| Literature DB >> 9059824 |
S G Vincent1, P R Cunningham, N L Stephens, A J Halayko, J T Fisher.
Abstract
In the present study we evaluated the performance of a software/scanner system that employed the Hewlett Packard (HP) ScanJet Plus and Scanplot Software for densitometric quantification of protein loads stained with Coomassie Brilliant Blue following sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE). Gels with bovine serum albumin (BSA) standards, ranging from 0.125 to 10 micrograms, were scanned using reflectance densitometry with 127 microns step size in both the x and y directions and a resolution of 200 dots per inch. Densitometric volume was calculated for each protein band from scanner output in the tagged image file format (TIFF) by a customized software package, Scanplot V. 4.05 (Cunningham Engineering). Protein loads between 0.125 and 10.0 micrograms vs. volume were fit by a second-order regression: Volume = -0.58 x protein load2 + 16.82 x protein load + 7.87 (r = 0.991, p < 0.01). The same gels were scanned and quantified using a transmittance laser densitometer; densitometric volumes measured by both systems were highly correlated (r2 = 0.981, p < 0.01). Additional gels of BSA, smooth muscle myosin heavy chain (myosin), and actin displayed linear relationships between protein loads up to 4.0 micrograms and densitometric volume reflecting unique dye binding properties. We conclude that accurate and reproducible quantitative densitometry of SDS-PAGE can be performed using the HP ScanJet Plus scanner and Scanplot software.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9059824 DOI: 10.1002/elps.1150180114
Source DB: PubMed Journal: Electrophoresis ISSN: 0173-0835 Impact factor: 3.535