Literature DB >> 9031494

Sensitivity and specificity of standard and rapid HIV-antibody tests evaluated by seroconversion and non-seroconversion low-titre panels.

E Kuun1, M Brashaw, A D Heyns.   

Abstract

BACKGROUND AND OBJECTIVES: The aim of this study is to compare the relative sensitivity and specificity of commercial HIV-antibody assays using seroconversion, non-seroconversion panels, and negative blood donor samples.
MATERIALS AND METHODS: We evaluated the sensitivity of five standard ELISA HIV-antibody assays: Vironostika HIV Uni-Form II, Abbott recombinant HIV-1/HIV-2 third-generation EIA, Biotest Anti-HIV-1/-2 recombinant, Recombigen HIV-1/ HIV-2 EIA and Wellcozyme HIV 1 + 2 (VK54/55), and three rapid screening tests, Capillus HIV-1/HIV-2, Abott Test Pack HIV-1/HIV-2 third-generation EIA, and Sensy-Test HIV 1/2. All tests were assessed using four panels of plasma samples obtained from individuals who were seroconverting and a low-titre HIV-antibody panel of samples. Specificity of the standard screening tests was determined on 3.500 HIV-antibody-negative blood donor samples.
RESULTS: There was no statistically significant difference in sensitivity between the five standard ELISA tests. One of these tests was significantly less specific than the others. The standard ELISA tests detected all the low-titre HIV-antibody-positive samples. Two of the rapid screening tests were significantly less sensitive on the seroconversion panels and all three tests failed to detect at least one of the positive samples in the low-titre panel.
CONCLUSIONS: The additional risk of using one or other of the standard ELISA tests under review of not detecting all HIV-positive units of blood is not statistically significant. Using some of the rapid screening tests will, however, add a significant additional risk. A rapid screening test should therefore be adopted only after careful consideration of the effect of a possible lack of sensitivity on the safety of the blood supply.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9031494     DOI: 10.1046/j.1423-0410.1997.00011.x

Source DB:  PubMed          Journal:  Vox Sang        ISSN: 0042-9007            Impact factor:   2.144


  7 in total

1.  HIV nucleic acid amplification testing versus rapid testing: it is worth the wait. Testing preferences of men who have sex with men.

Authors:  Joshua D O'Neal; Matthew R Golden; Bernard M Branson; Joanne D Stekler
Journal:  J Acquir Immune Defic Syndr       Date:  2012-08-01       Impact factor: 3.731

2.  Interval-censored data with misclassification: a Bayesian approach.

Authors:  Magda Carvalho Pires; Enrico Antônio Colosimo; Guilherme Augusto Veloso; Raquel de Souza Borges Ferreira
Journal:  J Appl Stat       Date:  2020-04-16       Impact factor: 1.416

3.  Evaluation of SAMBA II: A Qualitative and Semiquantitative HIV Point-of-Care Nucleic Acid Test.

Authors:  Lauren R Violette; Andy Cornelius-Hudson; Madison Snidarich; Lisa A Niemann; Sonny Michael Assennato; Allyson Ritchie; Neha Goel; Pollyanna R Chavez; Steven F Ethridge; David A Katz; Helen Lee; Kevin P Delaney; Joanne D Stekler
Journal:  J Acquir Immune Defic Syndr       Date:  2022-04-15       Impact factor: 3.771

4.  HIV testing in a high-incidence population: is antibody testing alone good enough?

Authors:  Joanne D Stekler; Paul D Swenson; Robert W Coombs; Joan Dragavon; Katherine K Thomas; Catherine A Brennan; Sushil G Devare; Robert W Wood; Matthew R Golden
Journal:  Clin Infect Dis       Date:  2009-08-01       Impact factor: 9.079

5.  Assessment of rapid tests for detection of human immunodeficiency virus-specific antibodies in recently infected individuals.

Authors:  Brian Louie; Ernest Wong; Jeffrey D Klausner; Sally Liska; Frederick Hecht; Terri Dowling; Martha Obeso; Susan S Phillips; Mark W Pandori
Journal:  J Clin Microbiol       Date:  2008-01-30       Impact factor: 5.948

6.  Incorporating validation subsets into discrete proportional hazards models for mismeasured outcomes.

Authors:  Amalia S Magaret
Journal:  Stat Med       Date:  2008-11-20       Impact factor: 2.373

7.  Relative accuracy of serum, whole blood, and oral fluid HIV tests among Seattle men who have sex with men.

Authors:  Joanne D Stekler; Joshua D O'Neal; Aric Lane; Fred Swanson; Janine Maenza; Claire E Stevens; Robert W Coombs; Joan A Dragavon; Paul D Swenson; Matthew R Golden; Bernard M Branson
Journal:  J Clin Virol       Date:  2013-12       Impact factor: 3.168

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.