Literature DB >> 9031310

Nd: YAG laser for Ahmed tube shunt occlusion by the posterior capsule.

Z Tessler1, S Jluchoded, G Rosenthal.   

Abstract

The authors describe the case of a 60-year-old aphakic woman who had suffered from glaucoma with uncontrolled intraocular pressure. She was treated by Ahmed tube shunt implantation. On the first postoperative day, the intraocular pressure was unacceptably high and the tip of the implant tube was seen to be embedded in a fold of the posterior capsule, with its opening covered by the posterior capsule. The authors used the Nd:YAG laser to cut open the fold of the posterior capsule that was covering the tip of the tube, thereby restoring the intraocular pressure to normal. To the best of the authors' knowledge, this is the first report of a tube shunt occlusion by the posterior capsule resolved by the Nd:YAG laser.

Entities:  

Mesh:

Year:  1997        PMID: 9031310

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers        ISSN: 1082-3069


  3 in total

1.  The outcome of the Ahmad Glaucoma Valve implantation for refractory glaucoma in Kuwait.

Authors:  Adel M Aljazzaf; Sidky M A Abdelmoaty; Abdelmutalib H Behbehani; Ahmad A Abdulmuez; Hadeel A Aljazzaf
Journal:  Saudi J Ophthalmol       Date:  2012-10-27

2.  Nd:YAG capsulotomy for Ahmed glaucoma drainage implant occlusion by the anterior capsule: a case report.

Authors:  Monica Kenney Ertel; Nathaniel Ryan Gelinas; Taylor John Slingsby; Leonard Keith Seibold; Malik Yaser Kahook; Jeffrey Raymond SooHoo
Journal:  BMC Ophthalmol       Date:  2021-02-06       Impact factor: 2.209

3.  Utility of Operative Glaucoma Tube Shunt Viscoelastic Bolus Flush.

Authors:  Sylvia L Groth; Kelsi L Greider; William Eric Sponsel
Journal:  J Curr Glaucoma Pract       Date:  2016-02-02
  3 in total