The importance of rigorous testing of circumference measuring callipers.

N J Dudley1, K Griffith.   

Abstract

On-screen circumference measurements are frequently made in obstetric ultrasound and are also made in other ultrasound specialties. The accuracy of electronic callipers is traditionally checked by linear measurement in a test object; this does not adequately test callipers used for nonlinear measurement. This technical note describes a simple test object designed for both linear and nonlinear calliper checks. Results show one machine to generate consistently large overmeasurements (15%). The clinical consequences of such overmeasurement may be serious. Appropriate quality control is essential.

Mesh:

Year:  1996        PMID: 9004436     DOI: 10.1016/s0301-5629(96)00100-7

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998