Entrance and exit dose measurements with semiconductors and thermoluminescent dosemeters: a comparison of methods and in vivo results.

T Loncol1, J L Greffe, S Vynckier, P Scalliet.   

Abstract

BACKGROUND AND
PURPOSE: In order to compare diodes and TLD for in vivo dosimetry, systematic measurements of entrance and exit doses were performed with semiconductor detectors and thermoluminescent dosemeters for brain and head and neck patients treated isocentrically with external photon beam therapy.
MATERIAL AND METHODS: Scanditronix EDP-20 diodes and 7LiF thermoluminescent chips, irradiated in a 8 MV linac, were studied with similar build-up cap geometries and materials in order to assure an equivalent electronic equilibrium. Identical calibration methodology was applied to both detectors for the dose determination in clinical conditions.
RESULTS: For the entrance dose evaluation over 249 field measurements, the ratio of the measured dose to the expected dose, calculated from tabulated tissue maximum ratios, was equal to 1.010 +/- 0.028 (1 s.d.) from diodes and 1.013 +/- 0.041 from thermoluminescent crystals. For the exit dose measurements, these ratios were equal to 0.998 +/- 0.049 and 1.016 +/- 0.070 for diodes and TLDs, respectively, after application of a simple inhomogeneity correction to the calculation of the expected exit dose.
CONCLUSIONS: Thermoluminescence and semiconductors led to identical results for entrance and exit dose evaluation but TLDs were characterised by a lower reproducibility inherent to the TL process itself and to the acquisition and annihilation procedures.

Entities:  

Mesh:

Year:  1996        PMID: 9004362     DOI: 10.1016/s0167-8140(96)01826-9

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  5 in total

1.  Is the in vivo dosimetry with the OneDosePlusTM system able to detect intra-fraction motion? A retrospective analysis of in vivo data from breast and prostate patients.

Authors:  Maria Daniela Falco; Marco D'Andrea; Alessia Lo Bosco; Mauro Rebuzzi; Elisabetta Ponti; Barbara Tolu; Grazia Tortorelli; Rosaria Barbarino; Luana Di Murro; Riccardo Santoni
Journal:  Radiat Oncol       Date:  2012-06-20       Impact factor: 3.481

2.  Evaluation of off-axis wedge correction factor using diode dosimeters for estimation of delivered dose in external radiotherapy.

Authors:  Mahmoud Allahverdi; Alireza Mohammadkarim; Mahbod Esfehani; Hasanali Nedaie; Alireza Shirazi; Ghazale Geraily
Journal:  J Med Phys       Date:  2012-01

3.  Entrance dose measurements for in-vivo diode dosimetry: Comparison of correction factors for two types of commercial silicon diode detectors.

Authors:  X R Zhu
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

4.  Patient skin dose measurements using a cable free system MOSFETs based in fluoroscopically guided percutaneous vertebroplasty, percutaneous disc decompression, radiofrequency medial branch neurolysis, and endovascular critical limb ischemia.

Authors:  Maria D Falco; Salvatore Masala; Matteo Stefanini; Roberto Fiori; Roberto Gandini; Paolo Bagalà; Daniele Morosetti; Eros Calabria; Alessia Tonnetti; Gianluca Verona-Rinati; Riccardo Santoni; Giovanni Simonetti
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

5.  Evaluation of target dose based on water-equivalent thickness in external beam radiotherapy.

Authors:  Behnaz Ghanbar Moghaddam; Masoud Vahabi-Moghaddam; Alireza Sadremomtaz
Journal:  J Med Phys       Date:  2013-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.