Strength recovery in fractured sheep tibia treated with a plate or an internal fixator: an experimental study with a two-year follow-up.

S Tepic1, A R Remiger, K Morikawa, M Predieri, S M Perren.   

Abstract

OBJECTIVES: Comparison of fracture healing with two different implants: a conventional Dynamic Compression Plate (DCP) and a new internal Point Contact Fixator (PC-Fix).
DESIGN: Randomized, prospective study in experimental animals. Observation times: 12, 24, 48 and 96 weeks, with six sheep per group.
SETTING: Following surgery, animals were kept with unrestricted weight-bearing in individual stalls for 12 weeks, thereafter in groups. ANIMALS: 56 adult Swiss mountain sheep. INTERVENTION: A standardized oblique fracture of the sheep tibia was reduced and compressed by a lag screw and "neutralized" with one of the implants. MAIN OUTCOME MEASUREMENTS: Standard radiographs were used for callus size measurements. After sacrifice the implant was removed and both the treated bone and the contralateral bone were tested for static strength in bending with the plate side under tension. Broken bones were processed for histological evaluation.
RESULTS: In the DCP group all six bones failed through the original fracture at 12 weeks. At 24 and 48 weeks two out of six, at 96 weeks one out of six bones failed through the original fracture, others through one of the screw holes. In the PC-Fix group there were no failures through the original fracture with a single exception at 96 weeks. The strength values in the PC-Fix groups of 12 and 96 weeks were significantly higher then in the corresponding DCP groups.
CONCLUSIONS: Healing of simple diaphysial fractures treated by PC-Fix was superior to that achieved by conventional plating. The histological evaluation suggested that the observed differences can be accounted for by the absence of implant-related cortical necrosis and by the circumferentially uninterrupted (if smaller) callus in the PC-Fix group.

Entities:  

Mesh:

Year:  1997        PMID: 8990027     DOI: 10.1097/00005131-199701000-00005

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  5 in total

1.  Net change in periosteal strain during stance shift loading after surgery correlates to rapid de novo bone generation in critically sized defects.

Authors:  Sarah H McBride; Scott Dolejs; Stefano Brianza; Ulf Knothe; Melissa L Knothe Tate
Journal:  Ann Biomed Eng       Date:  2011-01-27       Impact factor: 3.934

2.  [Optimizing the degree of fixation stability based on the strain theory].

Authors:  S M Perren
Journal:  Orthopade       Date:  2010-02       Impact factor: 1.087

3.  The dynamic locking screw (DLS) can increase interfragmentary motion on the near cortex of locked plating constructs by reducing the axial stiffness.

Authors:  Stefan Döbele; Carsten Horn; Stefan Eichhorn; Arne Buchholtz; Andreas Lenich; Rainer Burgkart; Andreas K Nüssler; Martin Lucke; Daniel Andermatt; Rudolf Koch; Ulrich Stöckle
Journal:  Langenbecks Arch Surg       Date:  2010-04-01       Impact factor: 3.445

4.  Clinical application of a minimally invasive cement-augmentable Schanz screw rod system to treat pelvic ring fractures.

Authors:  Paul Schmitz; Florian Baumann; Yves P Acklin; Boyko Gueorguiev; Michael Nerlich; Stephan Grechenig; Michael Bernd Müller
Journal:  Int Orthop       Date:  2018-05-21       Impact factor: 3.075

5.  Retrospective evaluation of perioperative and short term clinical outcomes in appendicular long bone skeleton fractures repaired via the string of pearls (SOP) locking plate system.

Authors:  Matthew R Field; Ryan Butler; Robert W Wills; Wilburn M Maxwell
Journal:  BMC Vet Res       Date:  2018-12-05       Impact factor: 2.741

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.