Literature DB >> 8983002

Effects of continuous-wave laser systems on stapes footplate.

S Jovanovic1, U Schönfeld, V Prapavat, A Berghaus, R Fischer, H Scherer, G J Müller.   

Abstract

BACKGROUND AND
OBJECTIVE: The aim of the present study was to clarify which of the presently available continuous-wave laser systems are best suited for application in stapes surgery. STUDY DESIGN/
MATERIALS AND METHODS: Isolated human stapes and bovine compact-bone platelets were used to investigate the connections between the parameters of various laser systems and their effects on bone tissue. The purpose was to optimize the laser parameters required to achieve a perforation measuring 500 microns to 600 microns in diameter. Three different laser systems were applied: the argon and CO2 laser in continuous wave (cw) mode and the CO2 laser in superpulse mode.
RESULTS: The suitability of the argon laser for stapedotomy is doubtful in view of the lower absorption coefficient of the stapes for the argon beam and the considerable influence which the degree of pigmentation of the irradiated medium exerts on its effect with the resultant poor reproducibility of the perforation diameter. The beam of the CO2 laser is far better absorbed at the footplate than that of the argon laser. This results in higher effectivity, lower thermic side effects, and better reproducibility of the perforation. The two modes of the CO2 laser do not show any appreciable differences.
CONCLUSION: The experimental results presented indicate that the CO2 laser in cw and superpulse mode is the most suitable of the systems now clinically applied in stapes surgery.

Entities:  

Mesh:

Year:  1996        PMID: 8983002     DOI: 10.1002/(SICI)1096-9101(1996)19:4<424::AID-LSM7>3.0.CO;2-U

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  8 in total

Review 1.  [Laser stapedotomy].

Authors:  A E Albers; W Wagner; K Stölzel; U Schönfeld; S Jovanovic
Journal:  HNO       Date:  2011-11       Impact factor: 1.284

2.  A comparison of hearing results following stapedotomy under local versus general anesthesia.

Authors:  Maureen Loewenthal; Nathan Jowett; Chia-Jung Busch; Rainald Knecht; Carsten V Dalchow
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-04-12       Impact factor: 2.503

3.  Photodynamic effect of argon and diode laser on cholesteatoma cell cultures after intravital staining with absorption enhancers.

Authors:  B Sedlmaier; A Franke; H Sudhoff; S Jovanovic; A Haisch
Journal:  Lasers Med Sci       Date:  2005-03-17       Impact factor: 3.161

4.  "One-shot" CO2 versus Er:YAG laser stapedotomy: is the outcome the same?

Authors:  Maria Raffaella Marchese; Alessandro Scorpecci; Francesca Cianfrone; Gaetano Paludetti
Journal:  Eur Arch Otorhinolaryngol       Date:  2010-10-19       Impact factor: 2.503

5.  Pressure and temperature changes in in vitro applications with the laser and their implications for middle ear surgery.

Authors:  Burkard Schwab; Georgios Kontorinis
Journal:  Int J Otolaryngol       Date:  2010-10-04

6.  ["One shot" CO2 laser stapedotomy].

Authors:  S Jovanovic; U Schönfeld; H Scherer
Journal:  HNO       Date:  2006-11       Impact factor: 1.284

7.  Long-term follow-up after "one-shot" CO2 laser stapedotomy: is the functional outcome stable during the years?

Authors:  Bruno Sergi; Daniela Lucidi; Eugenio De Corso; Gaetano Paludetti
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-03-23       Impact factor: 2.503

8.  CO2 laser stapedotomy safety: influence of laser energy and time on bone-conduction hearing levels.

Authors:  Uwe Schönfeld; Hu Weiming; Veit M Hofmann; Sergije Jovanovic; Andreas E Albers
Journal:  Eur Arch Otorhinolaryngol       Date:  2017-10-11       Impact factor: 2.503

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.