The efficient use of perimetry for neuro-ophthalmic diagnosis.

A Reitner1, M Tittl, E Ergun, R Baradaran-Dilmaghani.   

Abstract

AIMS/
BACKGROUND: Automated perimetry is both one of the most important diagnostic procedures in ophthalmology and the most difficult for the patient who is required to comply with strict and tiring conditions. This paper examines the use of a moving fixation target and a strategy for full threshold determination only in those data points found to be abnormal.
METHODS: 142 eyes in 71 patients were subjected to two types of visual field tests: the Dicon TKS 4000 Autoperimeter program 5 and the Humphrey field analyser program 30/2. The first procedure was compared with a commonly used instrument and strategy that is usually employed for the management of glaucoma patients.
RESULTS: Such a strategy was used in patients presenting with neuro-ophthalmic problems taking an average testing time of 14.9 minutes. The alternative test described here took an average of 3.9 minutes, without loss of diagnostic value.
CONCLUSION: The shorter testing time results in greater patient acceptance, fewer fatigue induced artefacts, and the possibility of completing a perimetry test with patients whose ability might seriously be challenged by a longer test. The reduced time required and the more natural condition of not having to fixate on an unmoving position were found not to reduce the diagnostic value of the data produced in patients presenting for various reasons at a neuro-ophthalmology clinic.

Entities:  

Mesh:

Year:  1996        PMID: 8976702      PMCID: PMC505646          DOI: 10.1136/bjo.80.10.903

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  2 in total

1.  Serial examination of the normal visual field using Octopus automated projection perimetry. Evidence for a learning effect.

Authors:  J M Wood; J M Wild; M K Hussey; S J Crews
Journal:  Acta Ophthalmol (Copenh)       Date:  1987-06

2.  Variability of the hill of vision and its significance in automated perimetry.

Authors:  N A Jacobs; I H Patterson
Journal:  Br J Ophthalmol       Date:  1985-11       Impact factor: 4.638

  2 in total
  3 in total

1.  Use of a portable head mounted perimetry system to assess bedside visual fields.

Authors:  D A Hollander; N J Volpe; M L Moster; G T Liu; L J Balcer; K D Judy; S L Galetta
Journal:  Br J Ophthalmol       Date:  2000-10       Impact factor: 4.638

2.  Comparative Study Between the SORS and Dynamic Strategy Visual Field Testing Methods on Glaucomatous and Healthy Subjects.

Authors:  Şerife Seda Kucur; Sebastian Häckel; Jan Stapelfeldt; Jeannine Odermatt; Milko E Iliev; Mathias Abegg; Raphael Sznitman; Rene Höhn
Journal:  Transl Vis Sci Technol       Date:  2020-12-02       Impact factor: 3.283

3.  Validating tablet perimetry against standard Humphrey Visual Field Analyzer for glaucoma screening in Indian population.

Authors:  Parul Ichhpujani; Sahil Thakur; Roopjit K Sahi; Suresh Kumar
Journal:  Indian J Ophthalmol       Date:  2021-01       Impact factor: 1.848

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.