Hemodynamic benefits of the Toronto Stentless Valve.

D F Del Rizzo1, B S Goldman, G T Christakis, T E David.   

Abstract

We report on 254 consecutive patients (170 male, 84 female) undergoing aortic valve replacement with the Toronto SPV Stentless Valve (St. Jude Medical, Inc., St. Paul, Minn.). Mean age (+/- standard deviation) was 62.1 +/- 11.6 years. Three patients (1%) received sizes 21 or 22 mm, 24 (9%) received size 23 mm, and 227 patients (89%) received sizes 25, 27, or 29 mm. Serial echocardiography was used to assess valve performance during a 3-year follow-up. Mean gradient decreased by 35.8% (p < 0.0001; 95% confidence interval -39.6%, -31.7%) from postoperative values to the 3- to 6-month follow-up and by 6.1% (p = 0.004; 95% confidence interval -10.1%, -2%) at each subsequent interval; effective orifice area increased by 17.2% (p = 0.0001; 95% confidence interval 12.0%, 22.6%) initially and by 4.4% (p < 0.001; 95% confidence interval 1.8%, 7.0%) thereafter. At 2 years of follow-up, mean gradient was 3.3 +/- 2.1 mm Hg and mean effective orifice area was 2.2 +/- 0.8 cm2. Studies on left ventricular mass were carried out on 84 patients. Left ventricular mass decreased by 14.3% (37.8 +/- 57.9 gm; p < 0.0001; 95% confidence interval -53.7, -21.9 gm) and left ventricular mass index decreased by 15.2% (21.1 +/- 30.5 gm/m2; p < 0.0001; 95% confidence interval -29.5, -12.7 gm/m2) from postoperative values to the 3- to 6-month follow-up interval. The reduction in residual gradient and potential regression in left ventricular hypertrophy may have a beneficial prognostic implication. We believe that the unique stentless design of the Toronto SPV Stentless Valve allows this to occur.

Entities:  

Mesh:

Year:  1996        PMID: 8975834     DOI: 10.1016/S0022-5223(96)70001-6

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  4 in total

1.  Dynamic obstruction, an unusual complication after aortic valve replacement with a stentless porcine valve.

Authors:  R P van Roosmalen; L H Baur; J Braun; D E Atsma; M G Hazekamp; E E van der Wall; H A Huysmans
Journal:  Int J Card Imaging       Date:  1999-06

2.  Improved postoperative outcomes with stentless aortic valve: a community hospital experience.

Authors:  S A Olenchock; J F Reed; A Brown; F M Garzia
Journal:  Heart       Date:  2003-05       Impact factor: 5.994

3.  Early experience of aortic valve replacement with the Freestyle stentless aortic bioprosthesis in elderly patients.

Authors:  S Ohtake; Y Sawa; T Sakaguchi; M Nishimura; H Satoh; G Matsumiya; Y Hayashi; H Matsuda
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2000-04