Comparison of continuous step-and-shoot versus step-and-shoot acquisition SPECT.

Z Cao1, C Maunoury, C C Chen, L E Holder.   

Abstract

UNLABELLED: This study compared the quality of SPECT images obtained with step-and-shoot mode (SSM) and continuous step-and-shoot mode (CSSM).
METHODS: Computer simulations of SSM and CSSM were performed with a two-dimensional Shepp-Logan head phantom and a high-resolution parallel-hole collimator. The effects of noise and photon attenuation were examined.
RESULTS: Without noise and without attenuation, small structures in reconstructed images obtained using fast CSSM (less than 4 sec per view plus the moving time from one view to the next) were slightly blurred and rotated in the opposite direction of detector motion. With both noise and attenuation, these artifacts were not visible, and the image quality obtained from CSSM, especially from fast CSSM, was improved as compared to that resulting from the corresponding SSM due to the increasing number of counts. The improvement of image quality became less significant with increasing acquisition time and in the presence of attenuation.
CONCLUSION: For fast SPECT, CSSM provided better image quality than the corresponding SSM, especially when attenuation effects were not present. For relatively long time SPECT without attenuation compensation that is typical for clinical studies, the image quality of CSSM was similar to that of SSM.

Mesh:

Year:  1996        PMID: 8970530

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  3 in total

1.  Recommendations for reducing radiation exposure in myocardial perfusion imaging.

Authors:  Manuel D Cerqueira; Kevin C Allman; Edward P Ficaro; Christopher L Hansen; Kenneth J Nichols; Randall C Thompson; William A Van Decker; Marko Yakovlevitch
Journal:  J Nucl Cardiol       Date:  2010-08       Impact factor: 5.952

2.  Instrumentation quality assurance and performance.

Authors:  Kenneth J Nichols; Stephen L Bacharach; Steven R Bergmann; S James Cullom; Edward P Ficaro; James R Galt; Gary V Heller; Jonathan Links; Josef Machac
Journal:  J Nucl Cardiol       Date:  2006-11       Impact factor: 5.952

3.  Verification of the effect of acquisition time for SwiftScan on quantitative bone single-photon emission computed tomography using an anthropomorphic phantom.

Authors:  Takuro Shiiba; Yuya Sekikawa; Shinji Tateoka; Nobutaka Shinohara; Yuuki Inoue; Yasuyoshi Kuroiwa; Takashi Tanaka; Yasushi Kihara; Takuroh Imamura
Journal:  EJNMMI Phys       Date:  2022-07-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.