Dual chamber pacing with a single lead system: initial clinical results.

J L Rey1, F el Ghelbazouri, C Tribouilloy.   

Abstract

A new mode of biphasic pacing was used in 26 patients to assess the feasibility of atrial pacing by means of the floating atrial ring electrodes of a single lead VDD permanent pacing system. During implantation, atrial pacing was possible in 25 patients with a 1-ms total pulse duration, a mean atrial threshold of 1.70 +/- 0.60 V (range, 0.6-3.0), and a mean diaphragmatic threshold of 6.7 +/- 2.5 V (range, 2.5-10.0). At 3 months, the atrial threshold had increased beyond 4.8 V in three patients. In the 22 other patients, the mean atrial threshold was 2.2 +/- 0.5 V (range, 1.50-3.50) in the supine position and 2.5 +/- 0.8 V (range, 1.5-4.8) in the sitting position. Stable atrial capture without diaphragmatic stimulation was achieved in 76% of patients.

Entities:  

Mesh:

Year:  1996        PMID: 8945039     DOI: 10.1111/j.1540-8159.1996.tb03223.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  The importance of right atrial pacing electrode position and pacing configuration for intra-atrial and inter-atrial conduction times.

Authors:  W M Hartung; D Hartung; H Saad; A Mittag; D Mahnkopf; H U Klein; R Willems
Journal:  J Interv Card Electrophysiol       Date:  2000-06       Impact factor: 1.900

2.  [Not Available].

Authors:  B Nowak; E Lorber; G Horstick; A Heimann; A Liebrich; S Rosocha; C Zellerhoff; O Kempski; E Himmrich; J Meyer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1998-02

Review 3.  The current status of single lead dual chamber sensing and pacing.

Authors:  H F Tse; C P Lau
Journal:  J Interv Card Electrophysiol       Date:  1998-09       Impact factor: 1.900

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.