Effect of surface roughness on hemolysis in a centrifugal blood pump.

Y Takami1, T Nakazawa, K Makinouchi, J Glueck, R Benkowski, Y Nosé.   

Abstract

Surface roughness of a blood pump is an important factor for blood cell damage. This study investigated the effect of surface roughness pertaining to hemolysis in a centrifugal pump. In vitro hemolysis tests were performed under cardiopulmonary bypass (CPB; 5 L/min, 350 mmHg) and left ventricular assist device (LVAD; 5 L/min, 100 mmHg) conditions using the pivot bearing supported Gyro centrifugal pump (C1E3). Seven types of pumps with impellers and housings with different surface roughness were prepared as follows: vapor polish (VP) housing and VP impeller; VP housing and sandpaper (SP) impeller; VP housing and fine sandblasting (FSB) impeller; VP housing and coarse sandblasting (CSB) impeller; SP housing and VP impeller; FSB housing and VP impeller; and CSB housing and VP impeller. The results revealed that 1) the effect of surface roughness on hemolysis was significantly larger with CPB than LVAD; 2) surface roughness, regardless of the impeller or housing, had little effect on hemolysis with LVAD; and 3) during CPB, the surface roughness of the pump housing had a larger effect on hemolysis than did that of the impeller. In conclusion, from a hemolytic point of view, it is likely that an extremely smooth pump housing is required for an impeller centrifugal pump for CPB. However, it is likely that a smooth surface is not as essential for this impeller centrifugal pump as for an LVAD.

Entities:  

Mesh:

Year:  1996        PMID: 8945006     DOI: 10.1097/00002480-199609000-00113

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  2 in total

1.  In vivo evaluation of the HeartWare centrifugal ventricular assist device.

Authors:  Egemen Tuzun; Katy Roberts; William E Cohn; Murat Sargin; Courtney J Gemmato; Branislav Radovancevic; O H Frazier
Journal:  Tex Heart Inst J       Date:  2007

2.  Properties of a monopivot centrifugal blood pump manufactured by 3D printing.

Authors:  Masahiro Nishida; Takumi Negishi; Daisuke Sakota; Ryo Kosaka; Osamu Maruyama; Toru Hyakutake; Katsuyuki Kuwana; Takashi Yamane
Journal:  J Artif Organs       Date:  2016-07-01       Impact factor: 1.731

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.