Influence of simulated altitude on the performance of five blood glucose meters.

J F Gautier1, A X Bigard, P Douce, A Duvallet, G Cathelineau.   

Abstract

OBJECTIVE: To determine the reliability of five blood glucose meters (BGMs) at various simulated altitudes using a hypobaric chamber. RESEARCH DESIGN AND METHODS: Blood glucose levels (ranged from 1.5 to 26.3 mmol/l, according to the reference method) were measured in 18 venous blood samples by each BGM at 200, 1,000, and every 500 m up to 4,000 m in a hypobaric chamber, where temperature and humidity were held constant.
RESULTS: Four BGMs underestimated and one overestimated blood glucose concentration while barometric pressure decreased. The average percent error varied in relation to simulated altitude from 0.26 +/- 4.8% (SD) at 200 m to -28.9 +/- 4.5% at 4,000 m (Glucometer 3; P < 0.05), from 28.4 +/- 5.7 to 49.3 +/- 5.9% (Accu-Chek Easy; P < 0.05), from -10.5 +/- 2.6 to 19.8 +/- 4.3% (Tracer, P < 0.05), from -5.5 +/- 2.6 to -11.2 +/- 3.0% (Reflolux; NS), and from 17.8 +/- 4.3 to 14.8 +/- 3.6% (One Touch; NS). The most accurate seemed to be the Reflolux, except for high blood glucose levels at simulated high altitudes. The One Touch II showed a good agreement, whatever the barometric pressure and the range of blood glucose concentrations. The highest underestimation was seen with the Glucometer 3.
CONCLUSIONS: Except for the Accu-Chek Easy, low barometric pressure underestimated the BGM results in comparison with measurements taken at simulated low altitudes. The lack of accuracy and consistency of performance > 2,000 m should be known by diabetic patients practicing sports activities, such as trekking or skiing at high altitudes.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8941478     DOI: 10.2337/diacare.19.12.1430

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  7 in total

1.  Diabetes and extreme altitude mountaineering.

Authors:  K Moore; C Thompson; R Hayes
Journal:  Br J Sports Med       Date:  2001-04       Impact factor: 13.800

Review 2.  Glucose homeostasis during short-term and prolonged exposure to high altitudes.

Authors:  Orison O Woolcott; Marilyn Ader; Richard N Bergman
Journal:  Endocr Rev       Date:  2015-02-12       Impact factor: 19.871

3.  Accuracy of handheld blood glucose meters at high altitude.

Authors:  Pieter de Mol; Hans G Krabbe; Suzanna T de Vries; Marion J Fokkert; Bert D Dikkeschei; Rienk Rienks; Karin M Bilo; Henk J G Bilo
Journal:  PLoS One       Date:  2010-11-12       Impact factor: 3.240

4.  Effects of simulated altitude on blood glucose meter performance: implications for in-flight blood glucose monitoring.

Authors:  Tolu Olateju; Joseph Begley; Daniel Flanagan; David Kerr
Journal:  J Diabetes Sci Technol       Date:  2012-07-01

Review 5.  Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge.

Authors:  Michael Erbach; Guido Freckmann; Rolf Hinzmann; Bernhard Kulzer; Ralph Ziegler; Lutz Heinemann; Oliver Schnell
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

6.  Study of the effect of altitude on the measurement of glycated haemoglobin using point-of-care instruments.

Authors:  Sandra W Veigne; Eugene Sobngwi; Brice E Nouthe; Joelle Sobngwi-Tambekou; Eric V Balti; Serge Limen; Mesmin Y Dehayem; Vicky Ama; Jean- Louis Nguewa; Maimouna Ndour-Mbaye; Alioune Camara; Naby M Balde; Jean- Claude Mbanya
Journal:  Cardiovasc J Afr       Date:  2015 Jan-Feb       Impact factor: 1.167

7.  Use of capillary blood glucose for screening for gestational diabetes mellitus in resource-constrained settings.

Authors:  Balaji Bhavadharini; Manni Mohanraj Mahalakshmi; Kumar Maheswari; Gunasekaran Kalaiyarasi; Ranjit Mohan Anjana; Mohan Deepa; Harish Ranjani; Miranda Priya; Ram Uma; Sriram Usha; Sonak D Pastakia; Belma Malanda; Anne Belton; Ranjit Unnikrishnan; Arivudainambi Kayal; Viswanathan Mohan
Journal:  Acta Diabetol       Date:  2015-04-28       Impact factor: 4.280

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.