Correction for respiration artifact in pulmonary blood pressure signals of ventilated patients.

S A Hoeksel1, J A Blom, J R Jansen, J J Schreuder.   

Abstract

OBJECTIVE: To develop an algorithm that corrects pulmonary artery pressure signals of ventilated patients for the respiration artifact. The algorithm should test the validity of the pulmonary pressure signal and differentiate between the cyclic respiration artifact and true measurement artifacts.
METHODS: The shape of each pulmonary pressure beat is described by eight characteristic features, including mean pressure value and the systolic and diastolic timing and pressure values. The features are corrected for the respiration artifact by fitting them in a least-squares sense on the first and second harmonics of the ventilator frequency. The corrected features are used by a signal validation algorithm, which adds a validity flag to each pressure beat. The validation algorithm rejects pressure beats with sudden changes in their shape but adapts itself when the changes persist.
RESULTS: The performance of the correction and validation technique was evaluated using pulmonary artery pressure signals of 30 patients who were scheduled for open heart surgery. The algorithm correctly recognized as invalid data those pressure signals disturbed by coagulation, surgical manipulations, or flushes of the pressure line. The algorithm marked on average 77 +/- 11% of the pulmonary pressure beats as valid.
CONCLUSIONS: The validation algorithm marked sufficient pressure beats as valid to update a trend display every 5 sec. The correction algorithm enabled the validation algorithm to differentiate between true measurement artifacts and the respiration artifact.

Entities:  

Mesh:

Year:  1996        PMID: 8934346     DOI: 10.1007/bf02077637

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  10 in total

Review 1.  The use of computers for controlling the delivery of anesthesia.

Authors:  D A O'Hara; D K Bogen; A Noordergraaf
Journal:  Anesthesiology       Date:  1992-09       Impact factor: 7.892

2.  Supervisory adaptive control of arterial pressure during cardiac surgery.

Authors:  J F Martin; N T Smith; M L Quinn; A M Schneider
Journal:  IEEE Trans Biomed Eng       Date:  1992-04       Impact factor: 4.538

3.  Automated recognition of corrupted arterial waveforms using neural network techniques.

Authors:  T Pike; R A Mustard
Journal:  Comput Biol Med       Date:  1992-05       Impact factor: 4.589

Review 4.  Computer control systems and pharmacological drug administration: a survey.

Authors:  D A Linkens; S S Hacisalihzade
Journal:  J Med Eng Technol       Date:  1990 Mar-Apr

5.  A general purpose neural network simulator system for medical data processing.

Authors:  T Odaka; H Agata; H Furutani; H Ogura
Journal:  J Med Syst       Date:  1994-12       Impact factor: 4.460

6.  bpshape wk4: a computer program that implements a physiological model for analyzing the shape of blood pressure waveforms.

Authors:  W C Ocasio; D R Rigney; K P Clark; R G Mark
Journal:  Comput Methods Programs Biomed       Date:  1993-04       Impact factor: 5.428

7.  Continuous, on-line, real-time spectral analysis of systemic arterial pressure signals.

Authors:  T B Kuo; S H Chan
Journal:  Am J Physiol       Date:  1993-06

8.  Interpretation of beat-to-beat blood pressure values in the presence of ventilatory changes.

Authors:  D M Ellis
Journal:  J Clin Monit       Date:  1985-01

9.  Multicenter trial of automated nitroprusside infusion for postoperative hypertension. Titrator Multicenter Study Group.

Authors:  W R Chitwood; D M Cosgrove; R M Lust
Journal:  Ann Thorac Surg       Date:  1992-09       Impact factor: 4.330

10.  Automatic control of arterial pressure after cardiac surgery. Evaluation of a microcomputer-based control system using glyceryl trinitrate and sodium nitroprusside.

Authors:  J R Colvin; G N Kenny
Journal:  Anaesthesia       Date:  1989-01       Impact factor: 6.955

  10 in total
  1 in total

1.  Detection of dicrotic notch in arterial pressure signals.

Authors:  S A Hoeksel; J R Jansen; J A Blom; J J Schreuder
Journal:  J Clin Monit       Date:  1997-09
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.