Literature DB >> 8902115

Transmission of Helicobacter pylori infection via flexible fiberoptic endoscopy.

T Akamatsu1, K Tabata, M Hironga, H Kawakami, M Uyeda.   

Abstract

BACKGROUND: Public concern has been raised with regard to the possibility of transmission of instrument mediated Helicobacter pylori infection after upper gastrointestinal endoscopy.
METHODS: Disinfection procedures for gastrointestinal endoscopes were surveyed in 20 Japanese institutions, and in vitro bactericidal activities of seven disinfectants against H. pylori were determined.
RESULTS: Screening tests for infection before endoscopy were not consistently performed; only 11 institutions always screened for hepatitis B virus, nine for hepatitis C virus, and two for tuberculosis. All 20 institutions used the same flexible fiberoptic endoscope on more than one patient in succession, with minimal cleanings only. Only two used glutaraldehyde for disinfection. Most used ethyl alcohol, benzalkonium chloride, or alkyldiaminoethylglycine hydrochloride as an external wipe. Bactericidal testing of nine strains of H. pylori against disinfectants revealed that ethyl alcohol (80%) and glutaraldehyde (0.5%) killed all nine strains within 15 seconds, whereas chlorhexidine gluconate (0.05%, 0.1%), benzalkonium chloride (0.025%, 0.1%), alkyldiaminoethylglycine hydrochloride (0.1%), povidone-iodine (0.1%), and sodium hypochlorite (150 ppm) killed all nine strains within 30 seconds.
CONCLUSION: H. pylori is readily killed by many common disinfectants and antiseptics. However, practices for disinfection of flexible fiberoptic endoscope were not appropriate.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8902115     DOI: 10.1016/s0196-6553(96)90028-0

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  3 in total

Review 1.  Helicobacter pylori.

Authors:  B E Dunn; H Cohen; M J Blaser
Journal:  Clin Microbiol Rev       Date:  1997-10       Impact factor: 26.132

Review 2.  Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.

Authors:  Julia Kovaleva; Frans T M Peters; Henny C van der Mei; John E Degener
Journal:  Clin Microbiol Rev       Date:  2013-04       Impact factor: 26.132

3.  Comparison of the efficacy of disinfectants in automated endoscope reprocessors for colonoscopes: tertiary amine compound (Sencron2®) versus ortho-phthalaldehyde (Cidex®OPA).

Authors:  Hyun Il Seo; Dae Sung Lee; Eun Mi Yoon; Min-Jung Kwon; Hyosoon Park; Yoon Suk Jung; Jung Ho Park; Chong Il Sohn; Dong Il Park
Journal:  Intest Res       Date:  2016-04-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.