| Literature DB >> 27175119 |
Hyun Il Seo1, Dae Sung Lee2, Eun Mi Yoon2, Min-Jung Kwon3, Hyosoon Park3, Yoon Suk Jung2, Jung Ho Park2, Chong Il Sohn2, Dong Il Park2.
Abstract
BACKGROUND/AIMS: To prevent the transmission of pathogens by endoscopes, following established reprocessing guidelines is critical. An ideal reprocessing step is simple, fast, and inexpensive. Here, we evaluated and compared the efficacy and safety of two disinfectants, a tertiary amine compound (TAC) and ortho-phthalaldehyde (OPA).Entities:
Keywords: Colonoscopes; Cost; Disinfectants; Endoscopes
Year: 2016 PMID: 27175119 PMCID: PMC4863052 DOI: 10.5217/ir.2016.14.2.178
Source DB: PubMed Journal: Intest Res ISSN: 1598-9100
Comparison of Sencron2® and Cidex®OPA
| Variable | Sencron2® | Cidex® OPA |
|---|---|---|
| Manufacturer | Bab Gencel Pharma & Chemical Ind. Co., Ankara, Turkey | Johnson & Johnson, New Brunswick, New Jersey, USA |
| Main ingredients | Tertiary amine compound | Ortho-phthalaldehyde |
| Exposure timea (min) | 5 | 5–129 |
| Price per bottleb | $ 143.00 (KRW 143,000)/L | $ 34.00 (KRW 34,000)/3.78 L |
| Form | Undiluted solution | Ready to use |
| Effective volume | 25 L (4%)c | 3.78 L |
| Price per single reprocessingb | $ 1.14 (KRW 1,144) | $ 1.79 (KRW 1,799) |
aRecommended exposure time at 20℃.
bApproximate price in Korea.
cDilute with tap water in an automated endoscope reprocessor.
KRW, Korean Won.
Positive Culture Rates After High-Level Disinfection With Orthophthalaldehyde (OPA) or Tertiary Amine Compound (TAC) Disinfectant
| Variable | TAC (Sencron2®) | OPA (Cidex® OPA) | |
|---|---|---|---|
| Suctiona | 0/50 | 1/50 | 1.000 |
| Biopsya | 2/50 | 5/50 | 0.436 |
| Operating channel | 1/50 | 0/50 | 1.000 |
| Total, n/N (%) | 3/150 (2) | 6/150 (4) | 0.501 |
aSuction and biopsy refer to the openings of the suction and biopsy channels, respectively.