Literature DB >> 8877472

Simulation studies promote technological development of radiofrequency phased array hyperthermia.

P Wust1, M Seebass, J Nadobny, P Deuflhard, G Mönich, R Felix.   

Abstract

A treatment planning program package for radiofrequency hyperthermia has been developed. It consists of software modules for processing three-dimensional computerized tomography (CT) data sets, manual segmentation, generation of tetrahedral grids, numerical calculation and optimisation of three-dimensional E field distributions using a volume surface integral equation algorithm as well as temperature distributions using an adaptive multilevel finite-elements code, and graphical tools for simultaneous representation of CT data and simulation results. Heat treatments are limited by hot spots in healthy tissues caused by E field maxima at electrical interfaces (bone/muscle). In order to reduce or avoid hot spots suitable objective functions are derived from power deposition patterns and temperature distributions, and are utilised to optimise antenna parameters (phases, amplitudes). The simulation and optimisation tools have been applied to estimate the improvements that could be reached by upgrades of the clinically used SIGMA-60 applicator (consisting of a single ring of four antenna pairs). The investigated upgrades are increased number of antennas and channels (triple-ring of 3 x 8 antennas and variation of antenna inclination. Significant improvement of index temperatures (1-2 degrees C) is achieved by upgrading the single ring to a triple ring with free phase selection for every antenna or antenna pair. Antenna amplitudes and inclinations proved as less important parameters.

Entities:  

Mesh:

Year:  1996        PMID: 8877472     DOI: 10.3109/02656739609023525

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  13 in total

Review 1.  Tumor targeting via EPR: Strategies to enhance patient responses.

Authors:  Susanne K Golombek; Jan-Niklas May; Benjamin Theek; Lia Appold; Natascha Drude; Fabian Kiessling; Twan Lammers
Journal:  Adv Drug Deliv Rev       Date:  2018-07-19       Impact factor: 15.470

2.  Effective learning strategies for real-time image-guided adaptive control of multiple-source hyperthermia applicators.

Authors:  Kung-Shan Cheng; Mark W Dewhirst; Paul R Stauffer; Shiva Das
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

3.  [A phase-I/II study on the local hyperthermia of cervical N2/N3 lymph node metastases].

Authors:  H Stahl; P Wust; R Graf; J Löffel; J Bier; H Riess; V Jahnke; R Felix
Journal:  Strahlenther Onkol       Date:  1997-04       Impact factor: 3.621

Review 4.  Heating technology for malignant tumors: a review.

Authors:  H Petra Kok; Erik N K Cressman; Wim Ceelen; Christopher L Brace; Robert Ivkov; Holger Grüll; Gail Ter Haar; Peter Wust; Johannes Crezee
Journal:  Int J Hyperthermia       Date:  2020       Impact factor: 3.914

5.  [Thermometry by measuring the chemical shift of lanthanide complex].

Authors:  P Konstanczak; P Wust; B Sander; S Schründer; T Frenzel; W Wlodarczyk; T Vogl; G Müller; R Felix
Journal:  Strahlenther Onkol       Date:  1997-02       Impact factor: 4.033

6.  The Kadota Fund International Forum 2004--clinical group consensus.

Authors:  J van der Zee; Z Vujaskovic; M Kondo; T Sugahara
Journal:  Int J Hyperthermia       Date:  2008-03       Impact factor: 3.914

7.  [Phase II study on preoperative radio-chemo-thermotherapy in locally advanced rectal carcinoma].

Authors:  B Rau; P Wust; J Gellermann; W Tilly; M Hünerbein; J Löffel; H Stahl; H Riess; V Budach; R Felix; P Schlag
Journal:  Strahlenther Onkol       Date:  1998-11       Impact factor: 3.621

8.  An approach to rapid calculation of temperature change in tissue using spatial filters to approximate effects of thermal conduction.

Authors:  Giuseppe Carluccio; Danilo Erricolo; Sukhoon Oh; Christopher M Collins
Journal:  IEEE Trans Biomed Eng       Date:  2013-01-22       Impact factor: 4.538

9.  Online feedback focusing algorithm for hyperthermia cancer treatment.

Authors:  Kung-Shan Cheng; Vadim Stakhursky; Paul Stauffer; Mark Dewhirst; Shiva K Das
Journal:  Int J Hyperthermia       Date:  2007-11       Impact factor: 3.914

10.  Design and evaluation of a hybrid radiofrequency applicator for magnetic resonance imaging and RF induced hyperthermia: electromagnetic field simulations up to 14.0 Tesla and proof-of-concept at 7.0 Tesla.

Authors:  Lukas Winter; Celal Özerdem; Werner Hoffmann; Davide Santoro; Alexander Müller; Helmar Waiczies; Reiner Seemann; Andreas Graessl; Peter Wust; Thoralf Niendorf
Journal:  PLoS One       Date:  2013-04-22       Impact factor: 3.240

View more