Verification and correction of setup deviations in tangential breast irradiation using EPID: gain versus workload.

J Pouliot1, A Lirette.   

Abstract

PURPOSE: When a deviation in the treatment setup is identified, when or how should it be corrected? With the aim of improving the precision and reducing the systematic errors while maintaining the workload to a minimum, a study was performed to define the proper set of actions for the correction of tangential breast setup deviations during the course of the treatment. PATIENTS AND METHODS: Clinical data were taken from a prospective study of more than 2200 on-line Electronic Portal Images (EPI) from 20 patients treated with tangential irradiation following a partial mastectomy for an early stage (I and II) breast cancer. The values of the central lung distance were then entered in the verification/correction procedure and modified to correct only the portion of the deviation most likely attributed to systematic errors according to the maximum likelihood. The verification/ correction procedure uses an action level equal to FML x alpha/square root of N, where alpha is proportional to the standard deviation (alpha = n sigma) and N is the number of consecutive fractions delivered after the start of the treatment. FML is the fraction of the deviation due to the systematic errors estimated from the maximum likelihood of the two distributions. In addition to n, one needs to specify Nmax, the maximum number of consecutive measurements without correction, to apply the procedure. The combination of these two parameters C(n,Nmax) will determine the reduction of systematic errors (gain) and the number of measurements and corrections (workload) associated to the procedure.
RESULTS: The effects of three combinations, C(1,1), C(2,2), and C(3,4) were studied. Also, the analysis of the results after application of the procedure with and without the factor of maximum likelihood made individually for each patient demonstrates the importance of the FML.
CONCLUSIONS: The verification/correction procedure with the inclusion of the FML can effectively improve the accuracy when applied to clinical data. With the specific workload related to measurements and corrections performed at our institution, we have found that in the case of the tangential breast treatment, the optimum combination uses an action level equal to 2 sigma and a maximum of two consecutive measurements without correction.

Entities:  

Mesh:

Year:  1996        PMID: 8873036     DOI: 10.1118/1.597722

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Correction of systematic set-up error in breast and head and neck irradiation through a no-action level (NAL) protocol.

Authors:  Eva M Lozano; Luis A Pérez; Javier Torres; Carmen Carrascosa; Miguel Sanz; Fermín Mendicote; Antonio Gil
Journal:  Clin Transl Oncol       Date:  2011-01       Impact factor: 3.405

2.  Dosimetric and isocentric variations due to patient setup errors in CT-based treatment planning for breast cancer by electronic portal imaging.

Authors:  Hari Mukundan; Deboleena Mukherjee; Kirti Tyagi; Sachin Taneja; Subhash Ranjan; Samaresh Sahu
Journal:  Med J Armed Forces India       Date:  2018-10-15

3.  Dosimetric Effects of the Interfraction Variations during Whole Breast Radiotherapy: A Prospective Study.

Authors:  Julian Jacob; Steve Heymann; Isabelle Borget; Isabelle Dumas; Elyes Riahi; Pierre Maroun; Patrick Ezra; Elena Roberti; Sofia Rivera; Eric Deutsch; Céline Bourgier
Journal:  Front Oncol       Date:  2015-09-16       Impact factor: 6.244

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.