Literature DB >> 8863115

Electrosurgery-induced ventricular fibrillation during pacemaker replacement--a unique mechanism.

A Aggarwal1, N E Farber, G S Kotter, M S Dhamee.   

Abstract

Arrhythmias and pacemaker malfunction are known to occur from the use of an electrosurgical device. The present case report describes a patient with sick sinus syndrome who experienced ventricular fibrillation while undergoing surgery. During replacement of his non-functioning cardiac pacemaker under general anesthesia, electrosurgery was used to ensure hemostasis. Electric current may have stimulated myocardial leads present in the surrounding tissue, leading to ventricular fibrillation. The patient was resuscitated from the episode without any residual sequelae. Microshock and possible mechanisms that can lead to ventricular arrhythmias in patients with pacemakers during electrosurgery are discussed.

Entities:  

Mesh:

Year:  1996        PMID: 8863115     DOI: 10.1007/bf02221756

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  11 in total

1.  EXTERNAL ELECTRIC STIMULATION OF THE HEART.

Authors:  P M ZOLL; A J LINENTHAL
Journal:  Ann N Y Acad Sci       Date:  1964-06-11       Impact factor: 5.691

2.  Surgical electrocautery and the runaway pacemaker syndrome.

Authors:  L I Heller
Journal:  Pacing Clin Electrophysiol       Date:  1990-09       Impact factor: 1.976

3.  The cause of stimulation with electrosurgical current.

Authors:  K S Foster; L A Geddes
Journal:  Med Instrum       Date:  1986 Nov-Dec

4.  Electrical thresholds for ventricular fibrillation in man.

Authors:  A B Watson; J S Wright; J Loughman
Journal:  Med J Aust       Date:  1973-06-16       Impact factor: 7.738

5.  Fibrillation resulting from pacemaker electrodes and electrocautery during surgery.

Authors:  J H Titel; A A el-Etr
Journal:  Anesthesiology       Date:  1968 Jul-Aug       Impact factor: 7.892

6.  Resetting of DDD pacemakers due to EMI.

Authors:  P H Belott; S Sands; J Warren
Journal:  Pacing Clin Electrophysiol       Date:  1984-03       Impact factor: 1.976

Review 7.  Pacemakers.

Authors:  J R Zaidan
Journal:  Anesthesiology       Date:  1984-04       Impact factor: 7.892

8.  Disturbances of heart rhythm produced by 50 Hz leakage currents in human subjects.

Authors:  E B Raftery; H L Green; M H Yacoub
Journal:  Cardiovasc Res       Date:  1975-03       Impact factor: 10.787

9.  Arrhythmias after the Mustard operation for transposition of the great arteries: a long-term study.

Authors:  C J Hayes; W M Gersony
Journal:  J Am Coll Cardiol       Date:  1986-01       Impact factor: 24.094

10.  Safety factor for precordial pacing: minimum current thresholds for pacing and for ventricular fibrillation by vulnerable-period stimulation.

Authors:  W D Voorhees; K S Foster; L A Geddes; C F Babbs
Journal:  Pacing Clin Electrophysiol       Date:  1984-05       Impact factor: 1.976

View more