| Literature DB >> 1700378 |
Abstract
A case of a unipolar surgical electrocautery-induced runaway pacemaker is described. The single chamber ventricular pacemaker began firing at 140 beats/min with intermittent capture and the abnormal rhythm persisted after cessation of the electrocautery. Emergent explanation of the pulse generator was required.Entities:
Mesh:
Year: 1990 PMID: 1700378 DOI: 10.1111/j.1540-8159.1990.tb02160.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976