Evaluation of K-space segmented cine sequences for fast functional cardiac imaging.

V Schülen1, F Schick, J Loichat, U Helber, P E Huppert, G Laub, C D Claussen.   

Abstract

RATIONALE AND
OBJECTIVES: In functional cardiac magnetic resonance imaging, reduction of measuring time is very important for many patients who are not able to rest motionless for long-lasting examinations. In this study, the image quality of sequences with k-space segmented data recording was compared with conventional gradient-echo sequences for cine imaging with flow compensation in applications on patients with normal and reduced ejection fractions.
METHODS: Thirty-one subjects (4 volunteers and 27 patients with cardiac diseases) were examined using different techniques for cine imaging of the left and right ventricles. The ejection fraction in the patients was calculated based on images of a conventional two-dimensional gradient-echo technique using a biplane ellipsoid model. The results from k-space segmented methods (3 to 9 Fourier lines per cardiac cycle for each phase image) were compared with the conventional images of the same short-axis view separately for groups of subjects with normal and reduced ejection fraction. The contrast between blood and myocardium at several sites of the heart and the homogeneity of the blood signal in the ventricle were evaluated for several phases of the heart cycle.
RESULTS: The segmentation in the acquisition of raw data allows reduction of measuring time to approximately 20% to 40% of the time required for conventional sequences in cine imaging of the heart. In patients with normal or only slightly reduced heart function (ejection fraction > or = 60%) the image quality of k-space segmented sequences was not significantly different from the conventionally recorded images. In contrast, patients with markedly lowered ejection fraction (< 60%) showed degraded results of the k-space segmented sequences compared with the conventional sequence (P < 0.001). The anterolateral border and the right ventricle especially were not sufficiently delineated by the k-space segmented sequences in these patients.
CONCLUSION: The k-space segmentation for the reduction of examination time is suitable for measuring heart volumes and functional parameters of patients expected to have a nearly normal ejection fraction, whereas for patients with markedly reduced cardiac function further technical improvements of segmented techniques are necessary.

Entities:  

Mesh:

Year:  1996        PMID: 8854198     DOI: 10.1097/00004424-199608000-00007

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  6 in total

Review 1.  Tools for cardiovascular magnetic resonance imaging.

Authors:  Ramkumar Krishnamurthy; Benjamin Cheong; Raja Muthupillai
Journal:  Cardiovasc Diagn Ther       Date:  2014-04

2.  Routine breath-hold gradient echo MRI-derived right ventricular mass, volumes and function: accuracy, reproducibility and coherence study.

Authors:  Farzin Beygui; Alain Furber; Stéphane Delépine; Gérard Helft; Jean-Philippe Metzger; Philippe Geslin; Jean Jacques Le Jeune
Journal:  Int J Cardiovasc Imaging       Date:  2004-12       Impact factor: 2.357

3.  Assessment of myocardial viability in a porcine model of chronic coronary artery stenosis with dual dose dobutamine magnetic resonance imaging.

Authors:  S Dymarkowski; M Szilard; A Maes; X S Liu; Y M Huang; E Verbeken; L Mortelmans; J Bogaert; I De Scheerder
Journal:  Int J Cardiovasc Imaging       Date:  2003-02       Impact factor: 2.357

4.  Assessment of Left Ventricular Systolic Function by Cardiovascular Magnetic Resonance Compressed Sensing Real-Time Cine Imaging Combined With Area-Length Method in Normal Sinus Rhythm and Atrial Fibrillation.

Authors:  Gang Yin; Chen Cui; Jing An; Kankan Zhao; Kai Yang; Shuang Li; Xinling Yang; Jiaxin Wang; Zhixiang Dong; Shiqin Yu; Jian He; Xiuyu Chen; Minjie Lu; Shihua Zhao
Journal:  Front Cardiovasc Med       Date:  2022-05-27

5.  Highly accelerated real-time cardiac cine MRI using k-t SPARSE-SENSE.

Authors:  Li Feng; Monvadi B Srichai; Ruth P Lim; Alexis Harrison; Wilson King; Ganesh Adluru; Edward V R Dibella; Daniel K Sodickson; Ricardo Otazo; Daniel Kim
Journal:  Magn Reson Med       Date:  2012-08-06       Impact factor: 4.668

6.  High spatial and temporal resolution cardiac cine MRI from retrospective reconstruction of data acquired in real time using motion correction and resorting.

Authors:  Peter Kellman; Christophe Chefd'hotel; Christine H Lorenz; Christine Mancini; Andrew E Arai; Elliot R McVeigh
Journal:  Magn Reson Med       Date:  2009-12       Impact factor: 4.668

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.