| Literature DB >> 8851246 |
G H Willemsen1, E J De Geus, C H Klaver, L J Van Doornen, D Carroll.
Abstract
The growing need for more advanced ambulatory monitoring has led to the development of an ambulatory monitor for impedance cardiography (VU-AMD). This paper presents two studies addressing the validity of the VU-AMD. In the first study, the cardiovascular responses of 25 subjects during various conditions were simultaneously recorded with the VU-AMD and a standard laboratory impedance device. Correlations between the responses of the ambulatory and laboratory devices were high, both inter- and intraindividually, except for stroke volume and cardiac output during exercise. In the second study, 26 subjects underwent 24-hr monitoring with the VU-AMD. The values obtained with the VU-AMD were realistic and varied in a predictable way over activity and posture. It is concluded that the VU-AMD is a valid device for the measurement of systolic time intervals in real-life situations, but its applicability for absolute stroke volume and cardiac output determination remains to be established.Entities:
Mesh:
Year: 1996 PMID: 8851246 DOI: 10.1111/j.1469-8986.1996.tb02122.x
Source DB: PubMed Journal: Psychophysiology ISSN: 0048-5772 Impact factor: 4.016