Examination technique, image quality, and patient dose in paediatric radiology. A survey including 19 Swedish hospitals.

A Almén1, M Lööf, S Mattsson.   

Abstract

PURPOSE: Investigation of examination technique, image quality, and absorbed dose to the patients in paediatric radiology.
MATERIAL AND METHODS: In total, 19 Swedish hospitals participated in the study. Using a questionnaire, the hospitals described their examination technique for the pelvis, urinary tract, colon, scoliosis, and lung. The image quality and patient dose were experimentally studied for the simulated pelvis examination of a 1-year-old child. This examination was carried out with a test object containing a contrast-detail phantom. TL dosimeters were used to determine the absorbed dose on the surface of the phantom, approximately corresponding to the absorbed dose on the surface of the patient.
RESULTS: Examination techniques varied considerably among the hospitals, the most striking difference concerning the film-screen sensitivity. Consequently, there was a variation in the absorbed dose on the surface of the phantom, from 0.09 mGy to 1.7 mGy (mean 0.65 mGy). For a large range of doses, 0.4-1.7 mGy, the image quality was not significantly different.
CONCLUSION: The unharmonized, and in many places unoptimized, examination techniques led to a great variation in the absorbed dose to the children examined.

Entities:  

Mesh:

Year:  1996        PMID: 8845265     DOI: 10.1177/02841851960371P171

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  4 in total

1.  The use of reference image criteria in X-ray diagnostics: an application for the optimisation of lumbar spine radiographs.

Authors:  A Almén; A Tingberg; J Besjakov; S Mattsson
Journal:  Eur Radiol       Date:  2004-04-01       Impact factor: 5.315

2.  Optimisation of scoliosis examinations in children.

Authors:  Jolanta Hansen; Anne Grethe Jurik; Bente Fiirgaard; Niels Egund
Journal:  Pediatr Radiol       Date:  2003-09-05

3.  Determination and Analysis of Film Reject Rate at Eight Selected Governmental Diagnostic X-Ray Facilities in Tigray Region, Northern Ethiopian.

Authors:  Messele Yisak Arbese; T Demissie Abebe; B Abraha Mesele
Journal:  J Med Phys       Date:  2018 Oct-Dec

4.  Analysis and Economic Implication of X-Ray Film Reject in Diagnostic Radiology Department of Jimma University Specialized Hospital, Southwest Ethiopia.

Authors:  Mesfin Zewdu; Elias Kadir; Melkamu Berhane
Journal:  Ethiop J Health Sci       Date:  2017-07
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.