BACKGROUND: We compared two T2-weighted turbo spin echo (TSE) sequences with a T2-weighted conventional SE (CSE) sequence to determine whether sequences derived from rapid acquisition with relaxation enhancement such as TSE could replace CSE for the detection and subsequent characterization of focal liver lesions.
METHODS: A total of 55 consecutive patients with 107 liver lesions underwent magnetic resonance imaging examinations at 1.5 Tesla, with a constant imaging protocol. TSE pulse sequences were acquired with eight echo trains (repetition time [TR], 4718 ms; echo time [TE], 90 ms; acquisition time [TA], 4.03 min; and a symmetric k-space ordering scheme) and 11 echo trains (TR, 4200 ms; TE, 140 ms; TA, 4.40 min; and an asymmetric k-space ordering scheme) and compared with CSE (TR, 2300 ms; TE, 45/90 ms; TA, 9.53 min). Images were analyzed qualitatively by scoring image quality and artifacts and counting focal liver lesions by independent reading with consensus obtained for discrepancies. Quantitative analysis was performed by measuring signal-to-noise (S/N), contrast-to-noise (C/N), and tumor-liver signal intensity (T/L) ratios.
RESULTS: T2-weighted TSE sequences provided better subjective image quality and reduced artifacts as compared with the T2-weighted CSE sequence. CSE and TSE sequences exhibited no statistically significant differences in liver S/N, lesion-liver C/N (CSE TE, 90 ms: 18.6 +/- 14.0; TSE TE, 90 ms: 16.5 +/- 12.9) and the detectability of focal liver lesions. Heavily T2-weighted TSE with a TE of 140 ms allowed correct characterization of focal liver lesions based on a T/L ratio of 3.0 in 84% of patients.
CONCLUSIONS: T2-weighted TSE sequences are as suited as CSE for the detection (TE, 90 ms), and appear to be superior for the characterization (TE, 140 ms), of focal hepatic lesions. Whether a single sequence, such as a double-echo TSE or a single-echo TSE sequence with a TE between 110 and 120 ms, might perform both functions as well or better than CSE is unknown. However, because of time savings, TSE eventually may be preferred over CSE.

Entities:  

Mesh:

Year:  1996        PMID: 8832864     DOI: 10.1007/s002619900097

Source DB:  PubMed          Journal:  Abdom Imaging        ISSN: 0942-8925


  21 in total

1.  The loss of small objects in variable TE imaging: implications for FSE, RARE, and EPI.

Authors:  R T Constable; J C Gore
Journal:  Magn Reson Med       Date:  1992-11       Impact factor: 4.668

Review 2.  First principles of fast spin echo.

Authors:  J Listerud; S Einstein; E Outwater; H Y Kressel
Journal:  Magn Reson Q       Date:  1992-12

3.  Magnetization transfer effects in multislice RARE sequences.

Authors:  P S Melki; R V Mulkern
Journal:  Magn Reson Med       Date:  1992-03       Impact factor: 4.668

4.  Fast spin-echo studies of contrast and small-lesion definition in a liver-metastasis phantom.

Authors:  D Chien; R V Mulkern
Journal:  J Magn Reson Imaging       Date:  1992 Jul-Aug       Impact factor: 4.813

5.  Measuring signal-to-noise ratios in MR imaging.

Authors:  L Kaufman; D M Kramer; L E Crooks; D A Ortendahl
Journal:  Radiology       Date:  1989-10       Impact factor: 11.105

6.  Differentiation of hepatic hemangiomas from metastases by dynamic contrast-enhanced MR imaging.

Authors:  B Hamm; E Fischer; M Taupitz
Journal:  J Comput Assist Tomogr       Date:  1990 Mar-Apr       Impact factor: 1.826

7.  Differentiation between small hepatic hemangiomas and metastases on MR images: importance of size-specific quantitative criteria.

Authors:  K Itoh; S Saini; P F Hahn; N Imam; J T Ferrucci
Journal:  AJR Am J Roentgenol       Date:  1990-07       Impact factor: 3.959

8.  T2-weighted MR imaging of focal hepatic lesions: comparison of various RARE and fat-suppressed spin-echo sequences.

Authors:  R C Semelka; J P Shoenut; R M Kroeker
Journal:  J Magn Reson Imaging       Date:  1993 Mar-Apr       Impact factor: 4.813

9.  Abdominal MR imaging: comparison of T2-weighted fast and conventional spin-echo, and contrast-enhanced fast multiplanar spoiled gradient-recalled imaging.

Authors:  R N Low; I R Francis; J S Sigeti; T K Foo
Journal:  Radiology       Date:  1993-03       Impact factor: 11.105

10.  T2-weighted MR imaging of the abdomen: fast spin-echo vs conventional spin-echo sequences.

Authors:  J V Catasca; S A Mirowitz
Journal:  AJR Am J Roentgenol       Date:  1994-01       Impact factor: 3.959

View more
  1 in total

1.  MRI characterization of 124 CT-indeterminate focal hepatic lesions: evaluation of clinical utility.

Authors:  Khaled M Elsayes; John R Leyendecker; Christine O Menias; Erica P Oliveira; Vamsidhar R Narra; William C Chapman; Moataz H Hassanien; Mohamed S Elsharkawy; Jeffrey J Brown
Journal:  HPB (Oxford)       Date:  2007       Impact factor: 3.647

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.