Selenium radiography versus storage phosphor and conventional radiography in the detection of simulated chest lesions.

C M Schaefer-Prokop1, M Prokop, A Schmidt, U Neitzel, M Galanski.   

Abstract

PURPOSE: To compare selenium detectors with three conventional and digital detector systems for the detection of simulated pulmonary lesions.
MATERIALS AND METHODS: Templates containing nodules, linear structures, and micronodular opacities were superimposed over an anthropomorphic chest phantom. The authors compared lesion detection with use of storage phosphor radiography (250 speed), selenium radiography (250 speed) with an antiscatter grid, selenium radiography (450 speed) without an antiscatter grid, an asymmetric screen-film system (400 speed), and a conventional screen-film system (250 speed). Detection performance of 10 radiologists was compared by using a multireader-multicase receiver operating characteristic analysis of variance.
RESULTS: For the detection of nodules, no statistically significant differences between imaging modes were seen. For the detection of micronodules and linear lesions, both selenium techniques were superior to all other modes (P < .05). In addition, the asymmetric screen-film radiographs were inferior (P < .05) to the conventional screen-film radiographs and to storage phosphor radiographs for the detection of micronodules.
CONCLUSION: The selenium detector improves detection of simulated fine linear and low-contrast micronodular details and appears to be superior to other detector systems for chest radiography.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8816519     DOI: 10.1148/radiology.201.1.8816519

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Comparison of liquid crystal versus cathode ray tube display for the detection of simulated chest lesions.

Authors:  Elisabeth Oschatz; Mathias Prokop; Martina Scharitzer; Michael Weber; Csilla Balassy; Cornelia Schaefer-Prokop
Journal:  Eur Radiol       Date:  2004-09-08       Impact factor: 5.315

2.  The use of reference image criteria in X-ray diagnostics: an application for the optimisation of lumbar spine radiographs.

Authors:  A Almén; A Tingberg; J Besjakov; S Mattsson
Journal:  Eur Radiol       Date:  2004-04-01       Impact factor: 5.315

3.  Detectability of simulated interstitial pneumonia on chest radiographs: comparison between irradiation side sampling indirect flat-panel detector and computed radiography.

Authors:  Y Yano; H Yabuuchi; A Kairada; N Tanaka; J Morishita; T Akasaka; Y Matsuo; T Kamitani; Y Yamasaki; M Nagao; M Sasaki
Journal:  Br J Radiol       Date:  2014-05-29       Impact factor: 3.039

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.