Polymeric surface modifications of tantalum stents.

A B Fontaine1, K Koelling, S D Passos, J Cearlock, R Hoffman, D G Spigos.   

Abstract

PURPOSE: To compare two kinds of polymer-coated tantalum stents with bare tantalum stents (control) to determine if the coatings can improve thromboresistance.
METHODS: Twenty-seven Fontaine-Dake stents were balloon expanded in three 8-mm x 80-cm polytetrafluoroethylene (PTFE) grafts; 9 stents were bare tantalum (T); 9 were coated with polyetherurethane (PL); and 9 were coated with parylene (PA). There were 9 stents placed in each graft as follows: 3 tantalum, 3 polyetherurethane, and 3 parylene. In swine whose platelets had been radiolabeled with indium 111, the ends of each stented graft were connected to 14F femoral and venous sheaths to create an ex vivo fistula. Each graft was exposed to blood for 30, 60, and 120 minutes. At the end of each test period, the stented grafts were disconnected from the sheaths, flushed with saline until clear, and then flushed with formalin. The stents were removed from the grafts, and a radionuclide well counter recorded radionuclide counts from each stent type at each period of blood contact. These values were converted to platelet density per 1000 microns 2. Stents were then photographed and scanned with electron microscopy (EM) for qualitative analysis. Possible significant differences in platelet adhesion with the three types of stents (both between stent groups and within stent groups) were examined using a two-tailed Student's t-test.
RESULTS: There were significantly fewer platelets adsorbed on PA versus T at all time periods (p < 0.005); on PL versus T at 60 and 120 minutes (p < 0.005); and on PA versus PL at 30 and 120 minutes (p < 0.0005). There was no significant difference in platelet density within each stent group (p = 0.1). Mean platelet density (number of platelets per 1000 microns 2 +/- SD) was as follows: at 30 minutes: T = 1891 +/- 965; PL = 373 +/- 193; and PA = 27 +/- 3; at 60 minutes: T = 6226 +/- 1621; PL = 1573 +/- 793; and PA = 1185 +/- 710; at 120 minutes: T = 5307 +/- 591; PL = 3164 +/- 318; and PA = 180 +/- 100. Gross inspection of the 120-minute groups demonstrated focal areas of thrombus on T, less on PL, and none on PA. Scanning EM demonstrated extensive platelet accumulation covering T at all time periods, less on PL, and even less on PA.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8800230     DOI: 10.1177/152660289600300306

Source DB:  PubMed          Journal:  J Endovasc Surg        ISSN: 1074-6218


  6 in total

1.  Biocompatibility of phosphorylcholine coated stents in normal porcine coronary arteries.

Authors:  D M Whelan; W J van der Giessen; S C Krabbendam; E A van Vliet; P D Verdouw; P W Serruys; H M van Beusekom
Journal:  Heart       Date:  2000-03       Impact factor: 5.994

2.  Cell response to a newly developed Ti-10Ta-10Nb alloy and its sputtered nanoscale coating.

Authors:  Young-Min Kim; Mong-Sook Vang; Hong-So Yang; Sang-Won Park; Hyun-Pil Lim
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

3.  A Micromachined Capacitive Pressure Sensor Using a Cavity-Less Structure with Bulk-Metal/Elastomer Layers and Its Wireless Telemetry Application.

Authors:  Kenichi Takahata; Yogesh B Gianchandani
Journal:  Sensors (Basel)       Date:  2008-04-02       Impact factor: 3.576

4.  Oligonucleotide and Parylene Surface Coating of Polystyrene and ePTFE for Improved Endothelial Cell Attachment and Hemocompatibility.

Authors:  Martina Schleicher; Jan Hansmann; Bentsian Elkin; Petra J Kluger; Simone Liebscher; Agnes J T Huber; Olaf Fritze; Christine Schille; Michaela Müller; Katja Schenke-Layland; Martina Seifert; Heike Walles; Hans-Peter Wendel; Ulrich A Stock
Journal:  Int J Biomater       Date:  2012-03-12

5.  Vapor sublimation and deposition to build porous particles and composites.

Authors:  Hsing-Ying Tung; Zhen-Yu Guan; Ting-Yu Liu; Hsien-Yeh Chen
Journal:  Nat Commun       Date:  2018-07-02       Impact factor: 14.919

6.  Biocompatibility of common implantable sensor materials in a tumor xenograft model.

Authors:  Mark E Gray; James Meehan; Ewen O Blair; Carol Ward; Simon P Langdon; Linda R Morrison; Jamie R K Marland; Andreas Tsiamis; Ian H Kunkler; Alan Murray; David Argyle
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2018-10-27       Impact factor: 3.368

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.