Multicenter experience with a pectoral unipolar implantable cardioverter-defibrillator. Active Can Investigators.

G H Bardy1, R Yee, W Jung.   

Abstract

OBJECTIVES: The purpose of this study was to prospectively examine in a multicenter study the methods of use, efficacy and complications of a unipolar cardioverter-defibrillator in patients at risk for sudden cardiac death.
BACKGROUND: Implantation of cardioverter-defibrillators in the pectoral region offers a significant opportunity to improve the management of patients with life-threatening arrhythmias. Unipolar, single-lead, pectoral implantable cardioverter-defibrillators might decrease related mortality, morbidity and costs in the care of such patients.
METHODS: From November 3, 1993 to May 8, 1995, a unipolar defibrillator (Medtronic model 7219C) was selected for use in 473 patients from 74 centers (386 [82%] men, 87 [18%] women; mean [+/- SD] age 59 +/- 13 years, range 16 to 88). The clinical indication for use was ventricular fibrillation in 157 patients, sustained ventricular tachycardia in 236, both ventricular tachycardia and ventricular fibrillation in 53 and syncope or inducible ventricular tachycardia/ventricular fibrillation in 27. Coronary artery disease was present in 323 patients (68%). The mean left ventricular ejection fraction was 0.36 +/- 0.15 (range 0.10 to 0.85). The distribution of New York Heart Association congestive heart failure was class I = 34%; class II = 45%; class III = 17%; and class IV = 2%.
RESULTS: The unipolar cardioverter-defibrillator was inserted successfully in 464 (98%) of 473 candidates. Effective defibrillation occurred with the first shock polarity tested in 88% of patients, after a polarity switch in 8% and after lead or generator repositioning in 2%. The stored energy defibrillation threshold was obtained at implantation in 339 patients (72%) and was 11.5 +/- 6.1 J, with 72% of patients having a defibrillation threshold < or = 12 J. The mean "skin-to-skin" implantation time was 96 +/- 45 min (range 25 to 335 min). Complications occurred in 29 patients (6%). Device therapy for 2,160 spontaneous ventricular tachycardia or fibrillation episodes occurred in 128 patients (27%) over a 2,732 device-month experience (range 0 to 17.2) and was effective in 98.7% of episodes. There were 14 deaths (10 nonsudden cardiac, 3 sudden cardiac, 1 noncardiac). Cumulative survival, on an intention-to-treat basis from all causes of death at 17.2 months, was 94.4%.
CONCLUSIONS: Unipolar pectoral implantable cardioverter-defibrillators can be inserted with a high likelihood of success in a relatively brief procedure. Defibrillation thresholds are low, morbidity is modest, and survival rates are good with this new type of implantable cardioverter-defibrillator.

Entities:  

Mesh:

Year:  1996        PMID: 8800117

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  8 in total

Review 1.  Dual- versus single-coil implantable defibrillator leads: review of the literature.

Authors:  Jörg Neuzner; Jörg Carlsson
Journal:  Clin Res Cardiol       Date:  2012-01-10       Impact factor: 5.460

2.  Experience with unipolar pectoral defibrillation.

Authors:  R K Reddy; G H Bardy
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

3.  Comparison of three different automatic defibrillator implantation approaches: pectoral implantation using conscious sedation reduces procedure times and cost.

Authors:  A Bollmann; N K Kanuru; D DeLurgio; P F Walter; J C Burnette; J J Langberg
Journal:  J Interv Card Electrophysiol       Date:  1997-11       Impact factor: 1.900

4.  Implantable defibrillators for life threatening ventricular arrhythmias. Are more effective than antiarrhythmic drugs in selected high risk patients.

Authors:  J P Causer; D T Connelly
Journal:  BMJ       Date:  1998-09-19

5.  Endovascular extraction techniques: Part 3: Results and indications in patients with an ICD.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-06       Impact factor: 2.380

6.  [Influence of waveform and configuration of electrodes on the defibrillation threshold of implantable cardioverter-defibrillators].

Authors:  M Block; D Hammel; G Breithardt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

7.  Complications and mortality of single versus dual chamber implantable cardioverter defibrillators.

Authors:  Ataallah Bagherzadeh; Zahra Emkanjoo; Majid Haghjoo; Maryam Moshkani Farahani; Abolfath Alizadeh; Mohammad Ali Sadr-Ameli
Journal:  Indian Pacing Electrophysiol J       Date:  2006-04-01

8.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.