Assessment of the optimal atrioventricular delay in dual chamber-paced patients by a portable scintillation probe (VEST).

L Mortelmans1, W Vanhecke, D Mertens, H Ector, C Timmermans, M De Roo, H De Geest, F Van de Werf.   

Abstract

BACKGROUND: The optimal atrioventricular delay in dual-chamber pacing differs from patient to patient. The availability of a portable scintillation probe (VEST) enables noninvasive monitoring of left ventricular function. METHODS AND
RESULTS: Hemodynamic variations were measured in 10 patients with programmable DDD pacemakers. The ejection fraction, stroke volume, and diastolic and systolic volume were evaluated, programming six different atrioventricular delays ranging from 75 to 200 msec, to determine the most favorable atrioventricular delay. Comparing left ventricular ejection fraction (LVEF), stroke volume, and end-diastolic and end-systolic volume at each DVI mode with a preceding DVI setting of 75 msec, all parameters at 200 msec were statistically different from those at 100 msec. An increase of LVEF and stroke volume and a decrease of end-systolic volume was found. In only five patients a switch of VVI mode to the optimal DVI mode results in an increase of LVEF of more than 5%.
CONCLUSIONS: Our study stresses the importance of optimizing atrioventricular delay. The VEST system permits these measurements, increasing the accuracy of the determination of optimal atrioventricular delay, and appears to be valuable in the management of patients with cardiac dual-chamber pacemakers.

Entities:  

Mesh:

Year:  1996        PMID: 8799251     DOI: 10.1016/s1071-3581(96)90092-6

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  27 in total

1.  Accuracy and reproducibility of left ventricular ejection fraction measurements using an ambulatory radionuclide left ventricular function monitor.

Authors:  L de Yang; C N Bairey; D S Berman; K J Nichols; T Odom-Maryon; A Rozanski
Journal:  J Nucl Med       Date:  1991-05       Impact factor: 10.057

2.  Myocardial demands of atrial-triggered versus fixed-rate ventricular pacing in patients with complete heart block.

Authors:  R Nordlander; S K Pehrsson; H Aström; J Karlsson
Journal:  Pacing Clin Electrophysiol       Date:  1987-09       Impact factor: 1.976

3.  Continuous monitoring of left ventricular function by an ambulatory radionuclide detector in patients with coronary artery disease.

Authors:  N Tamaki; T Yasuda; R H Moore; J B Gill; C A Boucher; A M Hutter; H K Gold; H W Strauss
Journal:  J Am Coll Cardiol       Date:  1988-09       Impact factor: 24.094

4.  The influence of left ventricular filling pressure on atrial contribution to cardiac output.

Authors:  B Greenberg; K Chatterjee; W W Parmley; J A Werner; A N Holly
Journal:  Am Heart J       Date:  1979-12       Impact factor: 4.749

5.  Assessment of the optimal atrio-ventricular delay in DDD paced patients by impedance plethysmography.

Authors:  M Eugene; G Lascault; R Frank; G Fontaine; Y Grosgogeat; A Teillac
Journal:  Eur Heart J       Date:  1989-03       Impact factor: 29.983

6.  Comparison of ambulatory left ventricular ejection fraction and blood pressure in systemic hypertension in patients with and without increased left ventricular mass.

Authors:  W M Breisblatt; C J Wolf; B McElhinny; R Salerni; V E Smith
Journal:  Am J Cardiol       Date:  1991-03-15       Impact factor: 2.778

7.  Evaluation of left ventricular diastolic function using an ambulatory radionuclide monitor: relationship to left ventricular systolic performance.

Authors:  M Ishibashi; T Yasuda; T P Rocco; N Alpert; R H Moore; H W Strauss
Journal:  Am Heart J       Date:  1990-07       Impact factor: 4.749

8.  Role of behavioral and psychological factors in mental stress-induced silent left ventricular dysfunction in coronary artery disease.

Authors:  M M Burg; D Jain; R Soufer; R D Kerns; B L Zaret
Journal:  J Am Coll Cardiol       Date:  1993-08       Impact factor: 24.094

9.  Usefulness of Doppler echocardiography in cardiac pacing: assessment of mitral regurgitation, peak aortic flow velocity and atrial capture.

Authors:  F T Zugibe; N C Nanda; S S Barold; T Akiyama
Journal:  Pacing Clin Electrophysiol       Date:  1983-11       Impact factor: 1.976

10.  The hemodynamic effects of ventricular pacing with and without atrioventricular synchrony in patients with normal and diminished left ventricular function.

Authors:  L A DiCarlo; F Morady; R B Krol; J M Baerman; M de Buitleir; M A Schork; S M Sereika; L Schurig
Journal:  Am Heart J       Date:  1987-10       Impact factor: 4.749

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.