Nonthoracotomy implantation of cardioverter defibrillators: preliminary experience with a defibrillation lead placed at the right ventricular outflow tract.

A S Tang1, P Hendry, W Goldstein, M S Green, M Luce.   

Abstract

UNLABELLED: Although morbidity and mortality associated with defibrillator implantation using a nonthoracotomy approach have decreased as compared with a thoracotomy approach, defibrillation thresholds have been higher and fewer patients satisfied implant criteria. It may be possible to improve on the success of nonthoracotomy defibrillator implantation by the placement of a right ventricular (RV) outflow defibrillation lead. Implantable cardioverter defibrillator implantation data of 30 consecutive patients with clinical VT or VF were reviewed. Three defibrillation leads were routinely used. When either pacing threshold at the RV apex was inadequate (n = 2) or 18-J shocks were not successful in terminating VF in 3 of 4 trials (n = 8), the RV apex lead was positioned to the RV outflow tract attaching to the septum. Defibrillation testing was first performed with the RV apex lead in combination with CS, SVC, and/or subcutaneous leads. Twenty patients satisfied implant criteria with a defibrillation threshold of 13.5 +/- 3.6 J. In 7 of the 10 patients, whose RV lead was repositioned to the RV outflow tract, this lead in combination with SVC, CS, or subcutaneous leads produced successful defibrillation at < or = 18 J or in 3 of 4 trials. This approach improved the overall success of nonthoracotomy implantation of defibrillators from 69% to 90%. After a follow-up of 27 +/- 6 months, there was no dislodgment of the RV outflow tract defibrillation leads.
CONCLUSIONS: This article reports the preliminary observation that placement of defibrillation leads to the RV outflow tract in humans was possible and without dislodgment. RV outflow tract offers an alternative for placement of defibrillation leads, which may improve on the success of nonthoracotomy defibrillator implantation.

Entities:  

Mesh:

Year:  1996        PMID: 8774827     DOI: 10.1111/j.1540-8159.1996.tb03393.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Endovascular extraction techniques: Part 3: Results and indications in patients with an ICD.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-06       Impact factor: 2.380

2.  A randomized comparison of defibrillation thresholds in the right ventricular outflow tract versus right ventricular apex.

Authors:  Michael Mollerus; Margaret Lipinski; Thomas Munger
Journal:  J Interv Card Electrophysiol       Date:  2008-05-10       Impact factor: 1.900

3.  Higher defibrillation threshold in methamphetamine cardiomyopathy patients with implantable cardioverter-defibrillator.

Authors:  Rakesh Malhotra; Shyam Patel; Tekchand Ramchand; Omar Al Nimri
Journal:  Indian Pacing Electrophysiol J       Date:  2017-07-08

4.  High defibrillation threshold: the science, signs and solutions.

Authors:  Sony Jacob; Victorio Pidlaoan; Jaspreet Singh; Aditya Bharadwaj; Mehul B Patel; Antonio Carrillo
Journal:  Indian Pacing Electrophysiol J       Date:  2010-01-07
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.