Evaluating data acquisition and smoothing functions of currently available videokeratoscopes.

M W Belin1, C D Ratliff.   

Abstract

PURPOSE: To compare the accuracy of computerized videokeratography systems using identical, calibrated test objects.
SETTING: Lions Eye Institute, Albany, New York.
METHODS: We evaluated the accuracy and smoothing of raw data acquisition (axial solution) of seven commercially available videokeratoscopes: Alcon EyeMap, Computed Anatomy TMS, EyeSys CAS, Humphrey MasterVue, Topcon CM-1000, Optikon Keratron, and TechnoMed C-Scan. We used six calibrated test objects to simulate clinical settings: spherical, spherocylindrical, simulated myopic ablation, hyperopic ablation, and a simulated central island.
RESULTS: None of the systems accurately imaged all objects. Although all systems imaged spherical objects with reasonable accuracy, errors greater than 4.0 diopters (D) frequently occurred in the central 6.0 mm optical zone (maximum error 10.0 D) Sources of error included excessive raw data smoothing, inability to read large transitions, loss of accuracy in the periphery, and poor central coverage.
CONCLUSION: The clinician should be aware of the potential limitations of corneal topography when making clinical decisions.

Entities:  

Mesh:

Year:  1996        PMID: 8733844     DOI: 10.1016/s0886-3350(96)80036-8

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  Wound healing anomalies after excimer laser photorefractive keratectomy: correlation of clinical outcomes, corneal topography, and confocal microscopy.

Authors:  R F Steinert
Journal:  Trans Am Ophthalmol Soc       Date:  1997

2.  The detection of keratoconus using novel metrics derived by anterior segment optical coherence tomography.

Authors:  Isa S K Mohammed; Sang Tran; Luis A Toledo-Espiett; Wuqaas M Munir
Journal:  Int Ophthalmol       Date:  2022-01-06       Impact factor: 2.029

3.  In vivo study of corneal responses to increased intraocular pressure loading.

Authors:  Ahmed Elsheikh; Charles W McMonnies; Charles Whitford; Gavin C Boneham
Journal:  Eye Vis (Lond)       Date:  2015-12-10

Review 4.  Scheimpflug imaging for keratoconus and ectatic disease.

Authors:  Michael W Belin; Renato Ambrósio
Journal:  Indian J Ophthalmol       Date:  2013-08       Impact factor: 1.848

5.  Inter-rater Reliability and Repeatability of Manual Anterior Segment Optical Coherence Tomography Image Grading in Keratoconus.

Authors:  Anna N Lin; Isa S K Mohammed; Wuqaas M Munir; Saleha Z Munir; Janet L Alexander
Journal:  Eye Contact Lens       Date:  2021-09-01       Impact factor: 3.152

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.