A cascade impactor entry port for MDI sprays with collection characteristics imitating a physical model of the human throat.

N C Miller1, A M Purrington.   

Abstract

PURPOSE: This work was performed in order to compare and contrast results obtained from cascade impactor measurements on metered dose inhalers (MDIs) using a variety of inlet ports.
METHODS: The collection characteristics of four cascade impactor ports (a physical model of the human throat, a simplified geometry intended to mimic the physical model, and two currently-used ports) were measured on a variety of MDI formulations.
RESULTS: The portion of the MDI spray which collects on the entry port depends in a complicated fashion on the characteristics of the formulation; in these studies the fraction of the total dose which was collected on the port ranged between about 20% and 90% of the total emitted dose. The collection characteristics of the simplified geometry closely corresponded to the physical model. The length of the flow path between the port and the impactor was varied, and found not to have a strong effect on the measured size distribution passing the port.
CONCLUSIONS: Ranking of various MDI formulations according to performance criteria as measured with a cascade impactor should be expected to depend on the particular inlet port which is used.

Entities:  

Mesh:

Year:  1996        PMID: 8692731     DOI: 10.1023/a:1016040509063

Source DB:  PubMed          Journal:  Pharm Res        ISSN: 0724-8741            Impact factor:   4.200


  6 in total

1.  The influence of sampling chamber dimensions on aerosol particle size measurement by cascade impactor and twin impinger.

Authors:  K Fults; T D Cyr; A J Hickey
Journal:  J Pharm Pharmacol       Date:  1991-10       Impact factor: 3.765

2.  Oropharyngeal deposition and delivery aspects of metered-dose inhaler aerosols.

Authors:  C S Kim; M A Eldridge; M A Sackner
Journal:  Am Rev Respir Dis       Date:  1987-01

3.  Small bore nozzle extensions to improve the delivery efficiency of drugs from metered dose inhalers: laboratory evaluation.

Authors:  R W Niven; R M Kacmarek; J D Brain; R A Peterfreund
Journal:  Am Rev Respir Dis       Date:  1993-06

4.  Effects of sampling chamber volume and geometry on aerodynamic size distributions of metered-dose inhalation aerosols measured with the Andersen cascade impactor.

Authors:  M Van Oort; R O Gollmar; R J Bohinski
Journal:  Pharm Res       Date:  1994-04       Impact factor: 4.200

5.  Simulated respiratory system for in vitro evaluation of two inhalation delivery systems using selected steroids.

Authors:  J J Sciarra; A Cutie
Journal:  J Pharm Sci       Date:  1978-10       Impact factor: 3.534

6.  Particle-size distribution of bitolterol mesylate solution delivered by four compressed-air nebulizer devices.

Authors:  W A Hunke; A B Yu
Journal:  Am J Hosp Pharm       Date:  1987-06
  6 in total
  1 in total

1.  Effect of oropharyngeal length in drug lung delivery via suspension pressurized metered dose inhalers.

Authors:  T Ehtezazi; D R Allanson; I D Jenkinson; C O'Callaghan
Journal:  Pharm Res       Date:  2006-05-26       Impact factor: 4.200

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.