Effects of dose reduction on digital chest imaging using a selenium detector: a study of detecting simulated diffuse interstitial pulmonary disease.

H P van Heesewijk1, Y van der Graaf, J C de Valois, M A Feldberg.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate the effect of dose reduction on the diagnostic performance of a new digital chest imaging system in which amorphous selenium is used as the X-ray detector. Diagnostic performance was assessed for the detection of simulated diffuse interstitial pulmonary disease (DIPD).
MATERIALS AND METHODS: DIPD was simulated by superimposing plastic sheets that contained small radiopaque objects (birdseed) on an anthropomorphic chest phantom. We varied the number of sheets from zero to four to simulate the degree of abnormality. We made 80 images with a standard X-ray dose, 80 images with 55% of the standard dose, and 80 images with 35% of the standard dose. Six observers were asked to indicate the presence of DIPD using a five-level scale of confidence. Two hundred forty chest images were then analyzed using receiver operating characteristic (ROC) curves.
RESULTS: The area under the ROC curve was 87.2 for all readers with standard-dose imaging (95% confidence interval [CI], 83.7-90.7), 91.7 with 55% of the standard dose (95% CI, 88.8-94.6), and 90.0 with the 35% dose (95% CI, 87.1-92.9). The area under the ROC curve for subtle DIPD (one superimposed sheet) was 75.3 for all readers with standard-dose imaging (95% CI, 67.1-83.5), 79.7 with 55% of the standard dose (95% CI, 71.9-87.5), and 70.3 with the 35% dose (95% CI, 61.7-78.9). For each dose, we observed a gradual improvement of the ROC curves with each additional sheet superimposed on the chest phantom (p < .001).
CONCLUSION: We found no significant difference in diagnostic performance among images made with standard X-ray dose, those made with a 55% dose, and those made with a 35% dose (95% CI).

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8686616     DOI: 10.2214/ajr.167.2.8686616

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  1 in total

1.  Image quality and dose differences caused by vendor-specific image processing of neonatal radiographs.

Authors:  William F Sensakovic; M Cody O'Dell; Haley Letter; Nathan Kohler; Baiywo Rop; Jane Cook; Gregory Logsdon; Laura Varich
Journal:  Pediatr Radiol       Date:  2016-08-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.