Literature DB >> 8685361

Aneurysm clip testing for ferromagnetic properties: clip variability issues.

E Kanal1, F G Shellock, J S Lewin.   

Abstract

To assess ferromagnetic properties of intracranial aneurysm clips reported to be nonferromagnetic, 1,765 Yasargil, 11 Sugita, and 15 Perneczky aneurysm clips were studied for rotation or translation on plate glass in a 1.5-T MR imager. Sixty-three clips (52 Yasargil, 11 Perneczky) weakly reoriented along the static magnetic resonance (MR) field. These results confirm the need for standardized testing for ferromagnetic properties for implantable metallic devices.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8685361     DOI: 10.1148/radiology.200.2.8685361

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  4 in total

1.  MR safety and imaging of neuroform stents at 3T.

Authors:  Arvind Nehra; Christopher J Moran; Dewitte T Cross; Colin P Derdeyn
Journal:  AJNR Am J Neuroradiol       Date:  2004-10       Impact factor: 3.825

2.  Clinical safety of cardiac magnetic resonance imaging at 3 T early after stent placement for acute myocardial infarction.

Authors:  Jérôme Jehl; Alexandre Comte; Sébastien Aubry; Nicolas Meneveau; François Schiele; Bruno Kastler
Journal:  Eur Radiol       Date:  2009-12       Impact factor: 5.315

3.  Aneurysm clips: evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems.

Authors:  Frank G Shellock; Jean A Tkach; Paul M Ruggieri; Thomas J Masaryk; Peter A Rasmussen
Journal:  AJNR Am J Neuroradiol       Date:  2003-03       Impact factor: 3.825

4.  In vitro evaluation of MR imaging issues at 3T for aneurysm clips made from MP35N: Findings and information applied to 155 additional aneurysm clips.

Authors:  F G Shellock; S Valencerina
Journal:  AJNR Am J Neuroradiol       Date:  2009-12-24       Impact factor: 3.825

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.