J J Kowalczyk1, J J Mulcahy. 1. Department of Urology, Indiana University Medical Center, Indianapolis, USA.
Abstract
PURPOSE: We reviewed our series of patients with an AMS700 Ultrex penile prosthesis whose erectile deformities were corrected by replacement with AMS700CX cylinders. MATERIALS AND METHODS: Ultrex cylinders were replaced by AMS700CX cylinders in 7 men with erectile deformities. RESULTS: Deformities were corrected successfully in all 7 patients. The penis is completely straight in 6 patients, while a mild 30-degree ventral curvature is present in 1. CONCLUSIONS: Erectile deformities with the AMS700 Ultrex penile prosthesis can be repaired successfully by replacement with the AMS700CX cylinders.
PURPOSE: We reviewed our series of patients with an AMS700 Ultrex penile prosthesis whose erectile deformities were corrected by replacement with AMS700CX cylinders. MATERIALS AND METHODS: Ultrex cylinders were replaced by AMS700CX cylinders in 7 men with erectile deformities. RESULTS: Deformities were corrected successfully in all 7 patients. The penis is completely straight in 6 patients, while a mild 30-degree ventral curvature is present in 1. CONCLUSIONS:Erectile deformities with the AMS700 Ultrex penile prosthesis can be repaired successfully by replacement with the AMS700CX cylinders.