A study of dose reduction using digital luminescence radiography for lateral skull radiography.

H Seifert1, R Kubale, T Hagen, B Kramann, H K Leetz.   

Abstract

For lateral skull radiography the minimum required radiation patient exposure to ensure adequate image quality was determined for digital luminescence radiography (DLR) in comparison with a screen-film system (speed class 200). Radiographs were produced with a grid technique on conventional X-ray equipment. A real prepared female head including a true fracture above the pars petrosa ossis temporalis was imaged. The tube current-time product (mAs), and thus the surface entrance dose, was varied systematically. Surface entrance dose was measured with TLD-100 rods. Image quality was judged by experienced radiologists according to the criteria: visual resolution, mean optical density, contrast and perceptibility of specific bone structures. Surface entrance dose was reduced from 0.46 to 0.20 mGy by application of DLR instead of speed class 200 screen-film system without loss of diagnostic information in clinical routine. This corresponds to a dose reduction potential of 57% showing a good agreement with the dose reduction potential of 52% obtained in a previous study using the Alderson head phantom.

Entities:  

Mesh:

Year:  1996        PMID: 8665130     DOI: 10.1259/0007-1285-69-820-311

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  1 in total

1.  Strict X-ray beam collimation for facial bones examination can increase lens exposure.

Authors:  R Powys; J Robinson; P L Kench; J Ryan; P C Brennan
Journal:  Br J Radiol       Date:  2012-02-28       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.