Sonographic signs of breast implant rupture.

L A Venta1, C G Salomon, M E Flisak, E R Venta, R Izquierdo, J Angelats.   

Abstract

OBJECTIVE: This study evaluated sonography as a screening test for breast implant rupture and developed diagnostic criteria for implant rupture. SUBJECTS AND METHODS: Women contemplating implant removal were evaluated prospectively with sonography. Implants were classified as normal, indeterminate, or ruptured. Individual sonographic signs were analyzed for their statistic association with implant rupture. To measure the degree of interobserver variation, static images were evaluated by two different observers.
RESULTS: Of 236 implants evaluated, surgical confirmation was available in 78, 22 of which were ruptured and 56, intact. Echogenic noise, multiple discontinuous parallel linear echoes, and echodense aggregates in the implant lumen were statistically associated with rupture. Based on sonographic findings, we classified 34 implants as intact (at surgery: 31 intact, 3 ruptured). We classified 19 as ruptured (at surgery: 11 ruptured, 8 intact). Of the 25 implants we classified as indeterminate, 17 were intact and 8 were ruptured at surgery. Therefore, sonography had a positive predictive value of 58%, a negative predictive value of 91%, a sensitivity of 50%, and a specificity of 55%. Receiver operating characteristic analysis suggests a learning curve effect and no significant interobserver variation.
CONCLUSIONS: A normal sonographic result is highly predictive of an intact implant. Thus, sonography is useful in evaluating symptomatic women or women concerned about implant rupture. An indeterminate sonographic result suggests the need for further testing.

Entities:  

Mesh:

Year:  1996        PMID: 8633455     DOI: 10.2214/ajr.166.6.8633455

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

Review 1.  Imaging findings of mammary and systemic silicone deposition secondary to breast implants.

Authors:  Naziya Samreen; Katrina N Glazebrook; Asha Bhatt; Sudhakar K Venkatesh; Brendan P McMenomy; Anupam Chandra; Shuai Leng; Kalie E Adler; Cynthia H McCollough
Journal:  Br J Radiol       Date:  2018-05-10       Impact factor: 3.039

Review 2.  The effect of study design biases on the diagnostic accuracy of magnetic resonance imaging for detecting silicone breast implant ruptures: a meta-analysis.

Authors:  Jae W Song; Hyungjin Myra Kim; Lillian T Bellfi; Kevin C Chung
Journal:  Plast Reconstr Surg       Date:  2011-03       Impact factor: 4.730

3.  Dual-energy CT for the evaluation of silicone breast implants.

Authors:  Thorsten R C Johnson; Isabelle Himsl; Karin Hellerhoff; Doris Mayr; Dorothea Rjosk-Dendorfer; Nina Ditsch; Bernhard Krauss; Klaus Friese; Maximilian F Reiser; Miriam S Lenhard
Journal:  Eur Radiol       Date:  2012-10-13       Impact factor: 5.315

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.