Abstract

Current routine methods for assessment of pulp vitality rely on stimulation of A-delta nerve fibers and give no direct indication of blood flow within the pulp. Recent papers have suggested that pulse oximeters may be used to diagnose pulp vitality by detection of blood flow. In this study, an optimized pulse oximeter probe for teeth was designed, built and tested using the Biox 3740 Oximeter (Ohmeda, Louisville, CO). Following preliminary in vitro tests, the probe was tested clinically. Pulse waveforms from maxillary and mandibular anterior teeth were noted. Simultaneous readings from the subjects' finger were used as controls. Pulse wave readings from the teeth were found to be synchronous with the finger probe, but not consistently. It was easier to maintain continuous readings from mandibular incisors than from maxillary incisors. The average percentage synchronization with the pulse was 28.95% for maxillary incisors and 50.28% for mandibular incisors. This difference was significant (p = 0.05). The overall accuracy of the commercial instrument was disappointing, and in its present form it was not considered to have clinical value.

Entities:  

Mesh:

Year:  1996        PMID: 8618088     DOI: 10.1016/S0099-2399(96)80283-4

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  4 in total

1.  Efficiency of new custom-made pulse oximeter sensor holder in assessment of actual pulp status.

Authors:  Krishnamachari Janani; P Ajitha; Raghu Sandhya; Haripriya Subbaiyan; Jerry Jose
Journal:  J Family Med Prim Care       Date:  2020-07-30

2.  Dental pulp testing: a review.

Authors:  Eugene Chen; Paul V Abbott
Journal:  Int J Dent       Date:  2009-11-12

Review 3.  Comparative Study of Pulp Vitality in Primary and Young Permanent Molars in Human Children with Pulse Oximeter and Electric Pulp Tester.

Authors:  Prinka Shahi; P B Sood; Arun Sharma; Manish Madan; Nishat Shahi; Geetanjali Gandhi
Journal:  Int J Clin Pediatr Dent       Date:  2015-08-11

4.  Pulp Vitality Testing with a Developed Universal Pulse Oximeter Probe Holder.

Authors:  Živilė Grabliauskienė; Roberta Zamaliauskienė; Greta Lodienė
Journal:  Medicina (Kaunas)       Date:  2021-01-23       Impact factor: 2.430

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.