Evaluating the accuracy of different sperm counting chambers by performing strict counts of photographed beads.

E Shiran1, J Stoller, Z Blumenfeld, P D Feigin, A Makler.   

Abstract

PURPOSE: To suggest a method for evaluating the accuracy of different kinds of sperm counting chambers by eliminating errors concerned with human skills or semen properties. In this method, various concentrations were prepared from stocks of commercially available latex beads (Accu-beads, Hamilton-Thorn Research, Beverly, MA). Samples from identical preparations were loaded into different types of chambers, namely, hemocytometer (Neubauer, improved double, Superior Ltd, Germany), Makler (Sefi Medical Instruments, Haifa, Israel) and Horwell (Horwell Ltd, London, UK). Beads were counted by both direct microscopic observations and by strict scanning of their photographed images.
RESULTS: In all cases, counts by direct observation were about 5% higher than strict counts of the same photographed beads. Counting photographed beads showed high reproducibility (average CV of 5.1%) between samples in the two wells of the hemocytometer. Counts of photographed beads, sampled from identical stocks, were on average slightly lower in the Makler chamber (20.7 x 10(6)/ml) and much higher in the Horwell chamber (47.4 x 10(6)/ml) than counts in the hemocytometer (21.5 x 10(6)/ml). Samples from three different batches of Accubeads revealed slight variation in counts between the batches and an average concentration of 11% above the number indicated on the commercial product. CONCLUSIONS; A technique that combined loading latex beads from identical stock into various chambers, proper covering of the tested samples and strict counting of photographed beads provided precise and reproducible results. By eliminating most errors related to human skills and semen properties, this method is suitable for evaluating the accuracy of counting chambers.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8574071     DOI: 10.1007/bf02211144

Source DB:  PubMed          Journal:  J Assist Reprod Genet        ISSN: 1058-0468            Impact factor:   3.412


  14 in total

1.  FACTORS AFFECTING HAEMOCYTOMETER COUNTS OF SPERM CONCENTRATION IN HUMAN SEMEN.

Authors:  M FREUND; B CAROL
Journal:  J Reprod Fertil       Date:  1964-10

2.  Comparison of computerized semen analysis with the conventional procedure in 322 patients.

Authors:  U A Knuth; E Nieschlag
Journal:  Fertil Steril       Date:  1988-05       Impact factor: 7.329

3.  The influence of chamber characteristics on the reliability of sperm concentration and movement measurements obtained by manual and videomicrographic analysis.

Authors:  K A Ginsburg; D R Armant
Journal:  Fertil Steril       Date:  1990-05       Impact factor: 7.329

4.  Automated semen analysis shows an increase in sperm concentration and motility with time in Makler chambers having excess sample volume.

Authors:  K Yanagida; K Hoshi; A Sato; L Burkman
Journal:  Hum Reprod       Date:  1990-02       Impact factor: 6.918

5.  [Use of Makler counting chamber in semen analysis].

Authors:  T Tomoyoshi; A Ishida; C J Kim
Journal:  Hinyokika Kiyo       Date:  1988-11

6.  Standardization and quality control of sperm concentration and sperm motility counts in semen analysis.

Authors:  D Mortimer; M A Shu; R Tan
Journal:  Hum Reprod       Date:  1986-08       Impact factor: 6.918

7.  Quality assurance for sperm concentration using latex beads.

Authors:  A J Peters; L J Zaneveld; R S Jeyendran
Journal:  Fertil Steril       Date:  1993-10       Impact factor: 7.329

8.  Discrepancies in sperm count using improved Neubauer, Makler, and Horwells counting chambers.

Authors:  G E Imade; O A Towobola; A S Sagay; J A Otubu
Journal:  Arch Androl       Date:  1993 Jul-Aug

9.  [A comparative study of the white cell pipette method and the Makler counting chamber method of counting spermatozoa].

Authors:  R Menkveld; J A Van Zyl; F S Stander; E Conradie; K Kopper
Journal:  S Afr Med J       Date:  1980-09-27

10.  The improved ten-micrometer chamber for rapid sperm count and motility evaluation.

Authors:  A Makler
Journal:  Fertil Steril       Date:  1980-03       Impact factor: 7.329

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.