| Literature DB >> 8572970 |
I Sakuma1, H Tadokoro, Y Fukui, T Dohi.
Abstract
Flow visualization is widely applied to evaluate rotary blood pumps; however, it is very difficult to visualize flow near the vanes of centrifugal blood pumps because the rotational speed of the impeller is usually several thousand rpm. In this study, a tracer method with a high speed video camera that can take more than 2,000 frames/s was utilized for flow visualization together with computer-assisted image measurement. This method visualized the complex secondary flow pattern near the vanes of the impeller, such as the vortex and recirculation. It also visualized the enhanced washout effect by the secondary washout vanes on the backside of the impeller. The proposed method was effective to analyze the flow pattern in the centrifugal blood pump by providing useful information for better design of the pump hemolysis and thrombus formation.Entities:
Mesh:
Year: 1995 PMID: 8572970 DOI: 10.1111/j.1525-1594.1995.tb02399.x
Source DB: PubMed Journal: Artif Organs ISSN: 0160-564X Impact factor: 3.094