Imaging 123I with a scintillation camera. A study of detection performance and quality factor concepts.

M S Bolmsjö, B R Persson, S E Strand.   

Abstract

Image quality, resolution and sensitivity of a scintillation camera equipped with various collimators have been investigated using high purity 123I. Pulse height distributions of 123I from a thyroid phantom partly in air and partly immersed in water demonstrate the substantial septa penetration of the 440 and 529 keV gamma rays of 123I with high resolution collimators. Line spread functions recorded first with the line source in air and then in water show that the area under the 'wings' is attributed mainly to septa penetration but with a marked contribution from scattering in water. The modulation transfer function evaluated from the line spread functions shows a sudden drop at low frequencies for high resolution collimators due to their high degree of septa penetration. The two concepts of 'figure of merit' also used are Qb=Sa [MTF]2, where Sa is the plane sensitivity and Qc=S2/(S+2B), where S is the true signal from the object and B is the total background or noise due to septa penetration and scatter. The image reproduction per unit time which is described by Qb is best for a high resolution converging collimator. The statistical accuracy per unit time which is used in dynamic studies is described by Qc and is best for a medium energy collimator.

Entities:  

Mesh:

Substances:

Year:  1977        PMID: 857263     DOI: 10.1088/0031-9155/22/2/006

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  Collimator choice in cardiac SPECT with I-123-labeled tracers.

Authors:  Yusuke Inoue; Ichiro Shirouzu; Toru Machida; Yasunori Yoshizawa; Fumihide Akita; Manabu Minami; Kuni Ohtomo
Journal:  J Nucl Cardiol       Date:  2004 Jul-Aug       Impact factor: 5.952

2.  Optimized acquisition and processing protocols for I-123 cardiac SPECT imaging.

Authors:  Ji Chen; Ernest V Garcia; James R Galt; Russell D Folks; Ignasi Carrio
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

3.  The dual photopeak-area method applied to scintillation camera measurements of effective depth and activity of in vivo 123I-distributions.

Authors:  S E Strand; B R Persson
Journal:  Eur J Nucl Med       Date:  1977-06-30

4.  Performance characteristics of a gamma camera over a wide range of energies.

Authors:  F R Hudson; J B Davis; A G Whittingham
Journal:  Eur J Nucl Med       Date:  1978-07-01

5.  Objective comparison of lesion detectability in low and medium-energy collimator iodine-123 mIBG images using a channelized Hotelling observer.

Authors:  Rebecca A Gregory; Iain Murray; Jonathan Gear; Matthew D Aldridge; Daniel Levine; Lucy Fowkes; Wendy A Waddington; Sue Chua; Glenn Flux
Journal:  Phys Med Biol       Date:  2016-12-14       Impact factor: 3.609

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.