| Literature DB >> 27973344 |
Rebecca A Gregory1, Iain Murray, Jonathan Gear, Matthew D Aldridge, Daniel Levine, Lucy Fowkes, Wendy A Waddington, Sue Chua, Glenn Flux.
Abstract
Iodine-123 mIBG imaging is widely regarded as a gold standard for diagnostic studies of neuroblastoma and adult neuroendocrine cancer although the optimal collimator for tumour imaging remains undetermined. Low-energy (LE) high-resolution (HR) collimators provide superior spatial resolution. However due to septal penetration of high-energy photons these provide poorer contrast than medium-energy (ME) general-purpose (GP) collimators. LEGP collimators improve count sensitivity. The aim of this study was to objectively compare the lesion detection efficiency of each collimator to determine the optimal collimator for diagnostic imaging. The septal penetration and sensitivity of each collimator was assessed. Planar images of the patient abdomen were simulated with static scans of a Liqui-Phil™ anthropomorphic phantom with lesion-shaped inserts, acquired with LE and ME collimators on 3 different manufacturers' gamma camera systems (Skylight (Philips), Intevo (Siemens) and Discovery (GE)). Two-hundred normal and 200 single-lesion abnormal images were created for each collimator. A channelized Hotelling observer (CHO) was developed and validated to score the images for the likelihood of an abnormality. The areas under receiver-operator characteristic (ROC) curves, Az, created from the scores were used to quantify lesion detectability. The CHO ROC curves for the LEHR collimators were inferior to the GP curves for all cameras. The LEHR collimators resulted in statistically significantly smaller Azs (p < 0.05), of on average 0.891 ± 0.004, than for the MEGP collimators, 0.933 ± 0.004. In conclusion, the reduced background provided by MEGP collimators improved 123I mIBG image lesion detectability over LEHR collimators that provided better spatial resolution.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27973344 PMCID: PMC6050520 DOI: 10.1088/1361-6560/62/1/17
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609
Figure 1.Static images of a 7 year old child with metastatic neuroblastoma deposits A acquired with a high-resolution collimator then B with a MEGP collimator.
Collimator characteristics with NEMA septal penetration and system sensitivity.
| Camera | Collimator | Hole size | Septal thickness | Length (mm) | Spatial resolution (mm) | Septal pen-etration % | System sensitivity (cps/MBq) ± error |
|---|---|---|---|---|---|---|---|
| Discovery (GE) | MEGP | 3.0 | 1.05 | 58.0 | 9.4 | 0.0 | 70.3 ± 1.2 |
| LEHR | 1.50 | 0.20 | 35.0 | 7.4 | 0.2 | 96.6 ± 1.7 | |
| Skylight (Philips) | MEGP | 2.95 | 1.14 | 48.0 | 11.3 | 0.0 | 101.1 ± 1.9 |
| LEHR | 1.40 | 0.15 | 32.8 | 7.4 | 1.1 | 109.3 ± 2.1 | |
| LEGP | 1.40 | 0.18 | 24.7 | 8.8 | 5.8 | 204.3 ± 3.9 | |
| Intevo (Siemens) | MELP | 2.94 | 1.14 | 40.64 | 12.5 | 0.0 | 115.6 ± 0.5 |
| LEHR | 1.11 | 0.16 | 24.05 | 7.5 | 16.3 | 185.9 ± 0.9 | |
| LEAP | 1.45 | 0.2 | 24.05 | 9.4 | 12.0 | 225.9 ± 1.0 | |
MEGP/LP = medium-energy general purpose/low penetration, LEG/AP = low-energy general/all purpose and LEHR = low-energy high-resolution.
Spatial resolution is supplied for collimated photons, as the FWHM at 10 cm from the collimator external surface in the manufacturers specification, for a 3/8″ crystal. The thicker (5/8″) Skylight crystal will provide poorer spatial resolution, however this value is not available from the manufacturer.
Error propagated from the uncertainty in the measured activity and square root of the region of interest counts, used in the NEMA calculation of sensitivity.
Figure 2.A cross-section of the abdomen-shaped phantom. The positions and activity concentrations of the fillable inserts are shown. These were scanned individually within the abdomen shaped shell alongside the spine insert. The shell dimensions vary within the range indicated along the axial length of the phantom.
Simulated activity concentration ranges in the resampled phantom images.
| Object | Simulated Range (kBq ml−1) |
|---|---|
| Abdomen | 0.53–0.79 |
| Liver | 1.6–2.4 |
| 3 cm sphere in liver | 36–143 |
| 2 cm sphere by spine | 23–202 |
| Vertebra | 1000–7000 |
Human observer scoring system.
| Score | Decision |
|---|---|
| 1 | Definitely normal |
| 2 | Probably normal |
| 3 | Equivocal |
| 4 | Maybe abnormal |
| 5 | Definitely ab normal |
Figure 3.An example of a high count image prior to Poisson resampling left, with a profile through the insert image overlaying the spine to mimic one active vetebra. The image on the right is the same geometry following Poisson resampling.
Figure 4.Example abnormal posterior images from the Skylight (Philips) gamma camera, corresponding to each of the available collimators.
Figure 5.CHO with a 0.008/cycle cut-off frequency versus the average human ROC curves for (a) the LEHR and (b) MEGP posterior images generated for the Philips Skylight gamma camera. The errors on the human curves are the standard deviation in the FPFs and TPFs.
CHO Validation (120 images for each collimator): areas under the ROC curves (figure 5) Az ± standard error.
| Collimator | Human | CHO |
|---|---|---|
| MEGP | 0.89 ± 0.11 | 0.93 ± 0.12 |
| LEHR | 0.92 ± 0.11 | 0.87 ± 0.12 |
Figure 6.The binormal fit to the ROC curves for each manufacturer and tested collimator.
CHO collimator comparison (400 images for each collimator): areas under the ROC curves (figure 6) Az ± standard error.
| Manufacturer | Collimator | |
|---|---|---|
| GE | MEGP | 0.932 ± 0.013 |
| LEHR | 0.894 ± 0.016 | |
| Philips | MEGP | 0.937 ± 0.012 |
| LEHR | 0.893 ± 0.016 | |
| LEGP | 0.903 ± 0.016 | |
| Siemens | MELP | 0.929 ± 0.015 |
| LEHR | 0.887 ± 0.017 | |
| LEAP | 0.920 ± 0.014 | |
Two tailed p-values of correlation between collimators Az.
| Manufacturer | Collimators | |
|---|---|---|
| GE | MEGP versus LEHR | 0.029 |
| Philips | MEGP versus LEHR | 0.012 |
| LEHR versus LEGP | 0.657 | |
| MEGP versus LEGP | 0.039 | |
| Siemens | MELP versus LEHR | 0.005 |
| LEHR versus LEAP | 0.011 | |
| MELP versus LEAP | 0.563 | |