Dosimetric considerations in radiation therapy of coin lesions of the lung.

E Yorke1, L Harisiadis, B Wessels, H Aghdam, R Altemus.   

Abstract

PURPOSE: The dose distribution in small lung lesions (coin lesions) is determined by the combined effects of reduced attenuation and electronic disequilibrium. The magnitude of the dose delivered also depends on the algorithm used to correct for reduced lung density. These effects are investigated experimentally and computationally for 10 MV photons. METHODS AND MATERIALS: Using a polystyrene miniphantom embedded in cork or cedar, thermoluminescent dosimetry and film dosimetry was performed to investigate interface effects and the central dose per monitor unit (MU). Three frequently applied calculation techniques--no density correction, ratio of tissue maximum ratios (TMRs), and the Batho correction--were also used to calculate the dose per MU. The measurements and calculations were compared with a one-dimensional phenomenological theory with parameters taken from the literature.
RESULTS: The measurements at the entrance surface and center of the miniphantom agreed well with the predictions of the phenomenological theory. The interface regions are usually thin enough (2-3 mm) to be clinically unimportant for 10 MV. Depending on the algorithm used to correct for decreased lung density, the lesion dose may be larger or smaller than the prescribed dose by as much as 20% in extreme cases. A clinical example is presented.
CONCLUSIONS: In comparing clinical results of treatments of small lung lesions, it is important to be aware of the density correction used.

Entities:  

Mesh:

Year:  1996        PMID: 8567352     DOI: 10.1016/0360-3016(95)02036-5

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  Accuracy Evaluation of EPL and ETAR Algorithms in the Treatment Planning Systems using CIRS Thorax Phantom.

Authors:  Mansour Zabihzadeh; Azizollah Rahimi; Hodjatollah Shahbazian; Sasan Razmjoo; Seyyed Rabie Mahdavi
Journal:  J Biomed Phys Eng       Date:  2021-08-01

2.  Dosimetric verification of radiotherapy treatment planning systems in Serbia: national audit.

Authors:  Laza Rutonjski; Borislava Petrović; Milutin Baucal; Milan Teodorović; Ozren Cudić; Eduard Gershkevitsh; Joanna Izewska
Journal:  Radiat Oncol       Date:  2012-09-12       Impact factor: 3.481

3.  An inhomogeneity correction algorithm for irregular fields of high-energy photon beams based on Clarkson integration and the 3D beam subtraction method.

Authors:  Sotirios Stathakis; Constantin Kappas; Kiki Theodorou; Nikos Papanikolaou; Jean-Claude Rosenwald
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

4.  Dosimetric comparison of extended dose range film with ionization measurements in water and lung equivalent heterogeneous media exposed to megavoltage photons.

Authors:  Paule M Charland; Indrin J Chetty; Shigeru Yokoyama; Benedick A Fraass
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.